FDA Adverse Event Injury Summary report: N

PRIMA

MDR report key: 18634273 · Received February 2, 2024

Report

Report Number
3005990499-2024-70194
Event Type
Injury
Date Received
February 2, 2024
Date of Event
February 26, 2024
Report Date
February 2, 2024
Manufacturer
KEYSTONE DENTAL, INC.
Product Code
DZE
UDI-DI
D76815620K0
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
DENTIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

KEYSTONE DENTAL INC. PROVIDES LABELING WITH ALL IMPLANTS. FAILURE TO OSSEOINTEGRATE AND LOSS OF OSSEOINTEGRATION (IMPLANT MOBILITY) IS IDENTIFIED AS AN ADVERSE REACTION AND IDENTIFIED IN THE LIST OF WARNINGS IN ALL ENDOSSEOUS DENTAL IMPLANT LABELING. THE SPECIFIC CAUSE FOR A PARTICULAR COMPLAINT IS OFTEN NOT READILY IDENTIFIED AS THERE ARE VARIOUS FACTORS THAT CONTRIBUTE TO THE RISK OF IMPLANT FAILURE. THESE INCLUDE PATIENT FACTORS SUCH AS PRIOR ORAL INFECTION, POOR BONE QUALITY OR QUANTITY, SYSTEMIC CONDITIONS SUCH AS DIABETES AND UNCONTROLLED HYPERTENSION. TECHNIQUE FACTORS SUCH AS OVERHEATING OF THE IMPLANT SITE MAY CAUSE NECROSIS OF THE BONE, PREVENTING GROWTH (PELAYO ET AL., 2008; TURKYILMAZ & AL., 2008). PATIENT HABITS SUCH AS TOBACCO USE (HEITZ-MAYFIELD & HUYNH-BA, 2009), ALCOHOL OR DRUG ABUSE, POOR ORAL HYGIENE, AND BRUXISM (SALVI & BRAEGGER, 2009) MAY ALSO LEAD TO IMPLANT FAILURE. IN ADDITION, IMPROPER SURGICAL TECHNIQUE CAN LEAD TO IMPLANT FAILURE AND/OR LOSS OF SUPPORTING BONE (SALVI & BRAEGGER, 2009).

Additional Manufacturer Narrative · 0

KEYSTONE DENTAL INC. PROVIDES LABELING WITH ALL IMPLANTS. FAILURE TO OSSEOINTEGRATE AND LOSS OF OSSEOINTEGRATION (IMPLANT MOBILITY) IS IDENTIFIED AS AN ADVERSE REACTION AND IDENTIFIED IN THE LIST OF WARNINGS IN ALL ENDOSSEOUS DENTAL IMPLANT LABELING. THE SPECIFIC CAUSE FOR A PARTICULAR COMPLAINT IS OFTEN NOT READILY IDENTIFIED AS THERE ARE VARIOUS FACTORS THAT CONTRIBUTE TO THE RISK OF IMPLANT FAILURE. THESE INCLUDE PATIENT FACTORS SUCH AS PRIOR ORAL INFECTION, POOR BONE QUALITY OR QUANTITY, SYSTEMIC CONDITIONS SUCH AS DIABETES AND UNCONTROLLED HYPERTENSION. TECHNIQUE FACTORS SUCH AS OVERHEATING OF THE IMPLANT SITE MAY CAUSE NECROSIS OF THE BONE, PREVENTING GROWTH (PELAYO ET AL., 2008; TURKYILMAZ & AL., 2008). PATIENT HABITS SUCH AS TOBACCO USE (HEITZ-MAYFIELD & HUYNH-BA, 2009), ALCOHOL OR DRUG ABUSE, POOR ORAL HYGIENE, AND BRUXISM (SALVI & BRAEGGER, 2009) MAY ALSO LEAD TO IMPLANT FAILURE. IN ADDITION, IMPROPER SURGICAL TECHNIQUE CAN LEAD TO IMPLANT FAILURE AND/OR LOSS OF SUPPORTING BONE (SALVI & BRAEGGER, 2009). UPDATING TO ADD HYPERTENSION

Description of Event or Problem · 0

LOSS OF OSSEOINTEGRATION.

Description of Event or Problem · 0

LOSS OF OSSEOINTEGRATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1347772 PRIMA PRIMACONNEX TC TAPERED RD 4.1X15 DZE KEYSTONE DENTAL, INC. 15620K WO-015260 D76815620K0

Patients

Seq Age Sex Outcome Treatment
1 58 YR Female Required Intervention