FDA Adverse Event Malfunction Summary report: N

BASAL/BOLUS INFUSOR 2ML X 2ML6 PACK

MDR report key: 1863411 · Received October 10, 2010

Report

Report Number
6000001-2010-03924
Event Type
Malfunction
Date Received
October 10, 2010
Date of Event
September 16, 2010
Report Date
September 16, 2010
Manufacturer
BAXTER HEALTHCARE - IRVINE
Product Code
FRN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE PHOTO OF THE DEVICE IS AVAILABLE FOR EVALUATION, PER THE CUSTOMER, HOWEVER, THE PHOTO HAS NOT YET BEEN RECEIVED. SHOULD THE PHOTO AND/OR ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED. THIS DEVICE IS MANUFACTURED FOR DISTRIBUTION OUTSIDE OF THE UNITED STATES (US); THEREFORE, IT DOES NOT CONTAIN A US 510K NUMBER. HOWEVER, THIS MDR IS BEING SUBMITTED BECAUSE IT IS THE SAME AS OR SIMILAR TO A PRODUCT DISTRIBUTED WITHIN THE US.

Additional Manufacturer Narrative · 1

(B)(4). ADDITIONAL NARRATIVE: DEVICE EVALUATION: ONE SAMPLE WAS RECEIVED BY BAXTER FOR EVALUATION. THE REPORTED CONDITION OF "RUPTURED RESERVOIR" WAS CONFIRMED. THIS IS A SINGLE USE DEVICE AND WILL NOT BE REPAIRED. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE WILL BE IDENTIFIED/ADDRESSED THROUGH THE CAPA INVESTIGATION, (B)(4). A BATCH REVIEW WAS CONDUCTED AND NO ISSUES WERE FOUND RELATED TO THE REPORTED CONDITION DURING THE MANUFACTURE OF THE LOT.

Description of Event or Problem · 1

IT WAS REPORTED TO BAXTER (B)(4) THAT THE RESERVOIR OF BASAL/BOLUS INFUSOR RUPTURED DURING FILLING. THE DEVICE WAS BEING FILLED WITH FENTANYL CITRATE AND NORMAL SALINE. AFTER THE DEVICE WAS FILLED WITH 90ML, THE RESERVOIR RUPTURED. THERE IS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION. NO ADDITIONAL INFORMATION IS AVAILABLE.