ASKU
Report
- Report Number
- 2183613-2010-00271
- Event Type
- Malfunction
- Date Received
- October 9, 2010
- Manufacturer
- MEDTRONIC MILACA INC.
- Product Code
- DTE
- PMA / PMN Number
- P820003/S72
- Removal / Correction Number
- Z-0430-2011
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4): ANALYSIS CONFIRMED THE DEVICE WOULD NOT POWER UP. THE CAUSE WAS THE MAIN PCB (PRINTED CIRCUIT BOARD). THE BATTERY CONTACTS WERE ALSO NOTED TO BE COMPRESSED, BATTERY DRAWER AND BATTERY DRAWER O-RING WERE CONTAMINATED, AND THE HEART LEAD FLEX WAS MISALIGNED.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE.
IT WAS REPORTED THE DEVICE WILL NOT POWER ON. NO INFORMATION WAS RECEIVED INDICATING PATIENT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ASKU | EXTERNAL PULSE GENERATOR | DTE | MEDTRONIC MILACA INC. | 5388 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |