FDA Adverse Event Malfunction Summary report: N

ASKU

MDR report key: 1863382 · Received October 9, 2010

Report

Report Number
2183613-2010-00271
Event Type
Malfunction
Date Received
October 9, 2010
Manufacturer
MEDTRONIC MILACA INC.
Product Code
DTE
PMA / PMN Number
P820003/S72
Removal / Correction Number
Z-0430-2011
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4): ANALYSIS CONFIRMED THE DEVICE WOULD NOT POWER UP. THE CAUSE WAS THE MAIN PCB (PRINTED CIRCUIT BOARD). THE BATTERY CONTACTS WERE ALSO NOTED TO BE COMPRESSED, BATTERY DRAWER AND BATTERY DRAWER O-RING WERE CONTAMINATED, AND THE HEART LEAD FLEX WAS MISALIGNED.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE.

Description of Event or Problem · 1

IT WAS REPORTED THE DEVICE WILL NOT POWER ON. NO INFORMATION WAS RECEIVED INDICATING PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ASKU EXTERNAL PULSE GENERATOR DTE MEDTRONIC MILACA INC. 5388 ASKU

Patients

Seq Age Sex Outcome Treatment
1 Other