FDA Adverse Event
Malfunction
Summary report: N
SPRINT QUATTRO SECURE
MDR report key: 1863372
·
Received October 9, 2010
Report
- Report Number
- 2649622-2010-11339
- Event Type
- Malfunction
- Date Received
- October 9, 2010
- Date of Event
- August 9, 2010
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- LWS
- PMA / PMN Number
- P920015/S24
- Removal / Correction Number
- ASKU
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE LEAD INTEGRITY ALERT (LIA) TRIPPED DUE TO AN SHORT INTERVAL COUNT (SIC) OF 632 SINCE (B)(6) 2010 AND NST EPISODES. THE EPISODES REVEALED T-WAVE OVERSENSING (TWOS). NO OBSERVABLE SENSING ISSUES CAN BE SEEN REAL-TIME AND THE IMPEDANCE TREND HAS REPORTEDLY BEEN STABLE. THE LEAD IS STILL IN USE. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPRINT QUATTRO SECURE | IMPLANTABLE TACHY LEAD | LWS | MEDTRONIC PUERTO RICO, INC. | 6947 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 54 YR | Other | 4194 IMPLANTABLE PACING LEAD| 5076 IMPLANTABLE PACING LEAD| D224TRK IMPLANTABLE PACEMAKER/CARDIO/DEFIB |