FDA Adverse Event
Injury
Summary report: N
ATTAIN ABILITY
MDR report key: 1863345
·
Received October 9, 2010
Report
- Report Number
- 2649622-2010-11390
- Event Type
- Injury
- Date Received
- October 9, 2010
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- OJX
- PMA / PMN Number
- P080006
- Removal / Correction Number
- ASKU
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) NO ANOMALIES FOUND, HOWEVER LEAD WAS STRETCHED, APPARENT EXPLANT DAMAGE NOTED, AND BLOOD NOTED ON ALL CONDUCTORS (NOT OBSTRUCTED); FULL LEAD RETURNED AND ANALYZED.
Description of Event or Problem · 1
IT WAS REPORTED THAT WHILE PLACING A NEW RIGHT VENTRICULAR PACE/SENSE LEAD THE LEFT VENTRICULAR LEAD DISLODGED. THIS LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ATTAIN ABILITY | IMPLANTABLE PACING LEAD | OJX | MEDTRONIC PUERTO RICO, INC. | 4196 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 66 YR | Required Intervention | 4076 IMPLANTABLE PACING LEAD| 6947 IMPLANTABLE TACHY LEAD| D224TRK IMPLANTABLE PACEMAKER/CARDIO/DEFIB |