UNICEL® DXC 600 PRO SYNCHRON® CLINICAL SYSTEM
Report
- Report Number
- 2050012-2010-00961
- Event Type
- Malfunction
- Date Received
- October 9, 2010
- Date of Event
- September 12, 2010
- Report Date
- October 8, 2010
- Manufacturer
- BECKMAN COULTER, INC.
- Product Code
- JJE
- PMA / PMN Number
- K042291
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- OTHER
Narratives
DURING TROUBLESHOOTING WITH CUSTOMER TECHNICAL SUPPORT (CTS), IT WAS DETERMINED THAT THE ISE (ION SELECTIVE ELECTRODE) DRAIN WAS OVERFLOWING. THE CTS INSTRUCTED THE CUSTOMER TO WEAR PERSONAL PROTECTIVE EQUIPMENT (PPE) INCLUDING EYE PROTECTION. THE CTS ALSO WARNED THE CUSTOMER THAT THE SPILLAGE IS HAZARDOUS AND THAT THEY SHOULD CONSULT MSDS AND FOLLOW LAB POLICY FOR DISPOSAL. THE CTS GUIDED THE CUSTOMER THROUGH DRAINING AND CLEANING ISE DRAIN CIRCUIT. THE CUSTOMER SUCCESSFULLY CLEANED AND PRIMED ALL ISE. AFTER 10 TIMES OF PRIMING ALL ISE, THE UNIT WOULD STILL DRAIN COMPLETELY. THE CUSTOMER CALIBRATED ALL ISE, RAN QC, AND PERFORMED PRECISION RUNS FOR ALL ISE CHEMISTRIES. THE SYSTEM WAS OPERATING ACCEPTABLY. AS OF (B)(4) 2010, THERE WERE NO FURTHER CALLS BY THE CUSTOMER, AND NO SERVICE VISIT WAS REQUIRED.
A CUSTOMER CONTACTED BECKMAN COULTER, INC (BCI) IN REGARDS TO CALIBRATION FAILURES ON UNICEL DXC 600 PRO SYNCHRON CLINICAL SYSTEM FOR CHLORIDE AND CALCIUM. THERE WAS NO EFFECT TO PATIENT OR TO USER. NO CHEMICAL EXPOSURE WAS NOTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL® DXC 600 PRO SYNCHRON® CLINICAL SYSTEM | CLINICAL CHEMSTRY ANALYZER | JJE | BECKMAN COULTER, INC. | DXC 600 PRO | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |