FDA Adverse Event
Malfunction
Summary report: N
ASKU
MDR report key: 1863255
·
Received October 9, 2010
Report
- Report Number
- 2183613-2010-00238
- Event Type
- Malfunction
- Date Received
- October 9, 2010
- Manufacturer
- MEDTRONIC MILACA INC.
- Product Code
- DTE
- PMA / PMN Number
- P820003/S72
- Removal / Correction Number
- ASKU
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) ANALYSIS CONFIRMED THE DEVICE WOULD NOT POWER UP. THE CAUSE WAS THE MAIN PCB (PRINTED CIRCUIT BOARD). THE LOWER CASE AND TWO SIDE BAIL COVERS WERE FOUND TO BE BROKEN AND BATTERY CONTACTS COMPRESSED.
Description of Event or Problem · 1
IT WAS REPORTED THE DEVICE WILL NOT WORK, EVEN WITH A NEW BATTERY. NO PATIENT INVOLVEMENT OR COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ASKU | EXTERNAL PULSE GENERATOR | DTE | MEDTRONIC MILACA INC. | 5388 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |