FDA Adverse Event Malfunction Summary report: N

CAPSURE SENSE

MDR report key: 1863225 · Received October 9, 2010

Report

Report Number
2649622-2010-10595
Event Type
Malfunction
Date Received
October 9, 2010
Date of Event
July 23, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P830061/S034
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) NO ANOMALIES FOUND. FULL LEAD RETURNED AND ANALYZED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR LEAD HAD EXCESSIVE NOISE ON THE EGM CHANNEL. THE IMPEDANCE WAS HIGH AND THERE WAS OVERSENSING WHEN PACING WAS EXPERIENCED. THE LEAD WAS ATTEMPTED BUT NOT IMPLANTED. ANOTHER LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSURE SENSE IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 4074 ASKU

Patients

Seq Age Sex Outcome Treatment
1 89 YR Other