SYVA® EMIT ® 2000 TACROLIMUS
Report
- Report Number
- 2517506-2024-00048
- Event Type
- Malfunction
- Date Received
- February 2, 2024
- Date of Event
- January 19, 2024
- Report Date
- April 15, 2024
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS INC.
- Product Code
- MLM
- UDI-DI
- 00842768001666
- PMA / PMN Number
- K060385
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
SIEMENS FILED THE INITIAL MDR 2517506-2024-00048 ON 02-FEB-2024. ADDITIONAL INFORMATION 21-MAR-2024: SIEMENS PERFORMED INITIAL TESTING AND REFERRED THE DATA FOR FURTHER REVIEW. ADDITIONAL INFORMATION 12-APR-2024: SIEMENS INVESTIGATED THIS ISSUE AND PROVIDED THE FOLLOWING CONCLUSION: THE EXPECTED BEHAVIOR OF THE SYVA® EMIT® 2000 TACROLIMUS ASSAY REAGENT IS AS FOLLOWS: INFORMATION SUMMARIZED BELOW CAN BE ATTAINED IN THE PRODUCT INSTRUCTIONS FOR USE (IFU), INSTRUMENT APPLICATION SHEET, AND VENDOR¿S CONTROL DOCUMENTATION. THE REAGENT IFU STATES THAT THE REAGENTS WILL REMAIN STABLE AFTER OPENING FOR 12 WEEKS OR UNTIL THE EXPIRATION DATE PRINTED ON THE LABEL (WHICHEVER IS SOONER), PENDING THAT THEY ARE STORED AT 2 ¿ 8°C, UPRIGHT, AND WITH CAPS TIGHTLY CLOSED. IFU STATES TO REFER TO APPLICATION SHEET FOR COMPLETE INSTRUCTIONS. APPLICATION SHEET STATES THAT REAGENTS MAY BE STORED ONBOARD, CAPPED WHEN NOT IN USE FOR UP TO 4 WEEKS OR AS QUALITY CONTROL (QC) RESULTS ARE WITHIN ACCEPTABLE LIMITS. QC EXPECTED RANGES ARE PROVIDED BY THE VENDOR, WITH THE EXPECTATION THAT INDIVIDUAL LABORATORIES WILL DEVELOP THEIR OWN QC RANGES USING THE EXPECTED RANGES AS A GUIDELINE. THE SYVA IFU ALSO PROVIDES INSTRUCTION FOR THE TESTING LABORATORY TO GENERATE THEIR OWN QC RANGES AND STATES THAT IN THE EVENT OF A QC FAILURE THE TESTING LABORATORY IS REQUIRED TO FOLLOW THE LABORATORY¿S QUALITY PROCEDURES. THE CUSTOMER IS ALLEGING INSTABILITY IN REAGENT OBSERVED BY QC RESULTS DRIFTING/FAILING LOW AND INCONSISTENT CLINICAL SAMPLE RESULTS. WHEN QC RESULTS DRIFT/FAIL LOW THE CUSTOMER IS REQUIRED TO RE-CALIBRATE AND TO PERFORM ADDITIONAL TROUBLESHOOTING. SINCE THE ATELLICA CH LOT 130034 CONTAINS THE SAME REAGENT USED IN THE SYVA EMIT 2000 TACROLIMUS REAGENT LOT S1 AND THE VIVA-E AND THE ATELLICA CH USE THE SAME PRINCIPLE OF SPECTROSCOPY TO MEASURE ANALYTE CONCENTRATION IN CALIBRATOR, CONTROL, OR PATIENT SAMPLE, SIEMENS PERFORMED TESTING USING THE ATELLICA CH REAGENT LOT 130034. THE PRINCIPLE OF SPECTROSCOPY USED IS BASED ON MEASURING THE ABSORBANCE OF LIGHT AT 340 DURING THE REACTION BETWEEN SAMPLE, REAGENT 1, AND REAGENT 2. SIEMENS' REVIEW OF THE ATELLICA CH DATA DID NOT CONFIRM THE OBSERVATIONS REPORTED BY THE CUSTOMER. THE SYVA EMIT 2000 TACROLIMUS ASSAY REAGENT HAS NO CALIBRATION INTERVAL CLAIMS AND THE VIVA-E APPLICATION SHEET INSTRUCTS THE OPERATOR TO ¿PREPARE A CALIBRATION CURVE WHENEVER A NEW LOT OF REAGENTS ARE USED OR AS INDICATED BY CONTROL RESULTS¿. THE PRODUCT PROBLEM IDENTIFIED BY THE CUSTOMERS IS NOT CONFIRMED. CUSTOMER SAMPLE HANDLING METHODS CANNOT BE RULED OUT AS A POTENTIAL RESULT IN ADVERSE ASSAY PERFORMANCE. THE SYVA EMIT 2000 TACROLIMUS ASSAY LOT S1 IS PERFORMING AS INTENDED. A PRODUCT PERFORMANCE ISSUE HAS NOT BEEN IDENTIFIED. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED. IN SECTION H6, THE INVESTIGATION FINDINGS AND INVESTIGATION CONCLUSIONS CODES WERE UPDATED.
AN OUTSIDE UNITED STATES (OUS) CUSTOMER CONTACTED A SIEMENS REMOTE SUPPORT CENTER (RSC) TO REPORT QUALITY CONTROL (QC) FAILURES AND ERRONEOUS PATIENT RESULTS WITH SYVA® EMIT® 2000 TACROLIMUS RESULTS THAT WERE OBTAINED ON VIVA-E® ANALYZER. THE FOLLOWING MAINTENANCE AND TROUBLESHOOTING STEPS WERE PERFORMED: CHECKED CUVETTES, DRY BLOCK, MIXER BEARING, BELT , VALVE KITS, MAINTENANCE DILUTER¿ REPLACED CUVETTE SAND DRY BLOCK, REPLACE WASH SOLUTION, REPLACE 0.1N HCL,10%NACLO N HCL,10%NACLO, DECONTAMINATION OF SYSTEM WITH MICRO CLEAN, DO CHECKR&CHECK SCV%<1MICRO CLEAN. CHECKR&CHECK SCV%<1. REPLACED THE REAGENT AND REPEATED TESTS. THE LIMITATIONS SECTION OF THE SYVA® EMIT® 2000 TACROLIMUS ASSAY INSTRUCTIONS FOR USE (IFU) STATES: "RESULTS OF THIS TEST SHOULD ALWAYS BE INTERPRETED IN CONJUNCTION WITH THE PATIENT¿S MEDICAL HISTORY, CLINICAL PRESENTATION AND OTHER FINDINGS." SIEMENS IS INVESTIGATING.
THE CUSTOMER REPORTED SYVA® EMIT ® 2000 TACROLIMUS RESULTS FOR MULTIPLE PATIENT SAMPLES ON THE VIVA-E® ANALYZER, THAT WERE CONSIDERED ERRONEOUS TO REPEAT TESTING ON THE SAME ANALYZER. ONLY RESULTS CONSIDERED CORRECT BY THE LABORATORY WERE REPORTED TO THE PHYSICIAN(S). ADDITIONALLY QUALITY CONTROL (QC) RECOVERED WITHIN RANGE AFTER REPROCESSING SEVERAL TIMES. PATIENT SAMPLES WERE NOT PROCESSED WHEN QC WAS OUT OF RANGE. PATIENT RESULTS WERE REPORTED ONLY WHEN QC WAS IN RANGE. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 232659 | SYVA® EMIT ® 2000 TACROLIMUS | ENZYME IMMUNOASSAY, TACROLIMUS | MLM | SIEMENS HEALTHCARE DIAGNOSTICS INC. | N/A | S1 | 00842768001666 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |