FDA Adverse Event Malfunction Summary report: N

CAPSUREFIX NOVUS

MDR report key: 1863138 · Received October 9, 2010

Report

Report Number
2649622-2010-10412
Event Type
Malfunction
Date Received
October 9, 2010
Date of Event
July 13, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039/S009
Removal / Correction Number
ASKU
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) HELIX/LOBE DISTORTED/BENT; FULL LEAD WAS RETURNED AND ANALYZED. BLOOD IN/ON HELIX/LOBE MECHANISM (SLEEVE HEAD).

Description of Event or Problem · 1

IT WAS REPORTED THAT "SCREW DIDN'T WORK PROPERLY UPON INSERTION" AND LOOKED TO BE UNCOILED UNDER FLUOROSCOPY. THE LEAD WAS NOT IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 5076 ASKU

Patients

Seq Age Sex Outcome Treatment
1 85 YR Required Intervention