FDA Adverse Event
Injury
Summary report: N
CAPSURE Z NOVUS
MDR report key: 1862919
·
Received October 9, 2010
Report
- Report Number
- 2649622-2010-10458
- Event Type
- Injury
- Date Received
- October 9, 2010
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- DTB
- PMA / PMN Number
- P850089/S38
- Removal / Correction Number
- ASKU
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THIS REPORT IS BASED SOLELY ON DEVICE ANALYSIS. THE EVENT OCCURRED OUTSIDE THE US. NO INFORMATION TO SUGGEST A DEVICE-RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4): OUTER INSULATION BREACHED DEPRESSION, HOWEVER BLOOD WAS NOTED ON DISTAL CONDUCTOR (NOT OBSTRUCTED); PROXIMAL SEGMENT RETURNED AND ANALYZED.
Description of Event or Problem · 1
THE LEAD WAS RETURNED TO THE MANUFACTURER, ANALYZED, AND TESTED OUT OF SPECIFICATION. FOLLOW-UP DID NOT YIELD ANY ADDITIONAL RELEVANT INFORMATION REGARDING THIS EVENT. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CAPSURE Z NOVUS | IMPLANTABLE PACING LEAD | DTB | MEDTRONIC PUERTO RICO, INC. | 5054 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other | 5554 IMPLANTABLE PACING LEAD| (B)(4) IMPLANTABLE PULSE GENERATOR |