FDA Adverse Event
Malfunction
Summary report: N
MAXIMO II VR
MDR report key: 1862868
·
Received October 9, 2010
Report
- Report Number
- 6000144-2010-05043
- Event Type
- Malfunction
- Date Received
- October 9, 2010
- Manufacturer
- MEDTRONIC MED REL, INC.
- Product Code
- LWS
- PMA / PMN Number
- P980016/S114
- Removal / Correction Number
- ASKU
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE DEVICE INDICATED AN EPISODE WAS A MONITORED VENTRICULAR TACHYCARDIA (VT), WHEN IT SHOULD HAVE BEEN A SUPRAVENTRICULAR TACHYCARDIA (SVT), DUE TO WAVELET NOT DISCRIMINATING APPROPRIATELY. THE DEVICE REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MAXIMO II VR | IMPLANTABLE PACEMAKER/CARDIO/DEFIB | LWS | MEDTRONIC MED REL, INC. | D284VRC | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 59 YR | Required Intervention | 6947 IMPLANTABLE TACHY LEAD |