FDA Adverse Event
Malfunction
Summary report: N
CAPSUREFIX
MDR report key: 1862849
·
Received October 9, 2010
Report
- Report Number
- 2649622-2010-10512
- Event Type
- Malfunction
- Date Received
- October 9, 2010
- Date of Event
- July 20, 2010
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- DTB
- PMA / PMN Number
- P930039/S2
- Removal / Correction Number
- ASKU
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE LEAD WARNING WAS TRIGGERED, AND THE PACING IMPEDANCE OF THE PATIENT'S BIPOLAR ATRIAL LEAD HAD DECREASED FROM 700 OHMS TO 344 OHMS AND THAT THE UNIPOLAR IMPEDANCE WAS HIGHER THAN THE BIPOLAR IMPEDANCE. THE LEAD IS STILL IN USE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CAPSUREFIX | IMPLANTABLE PACING LEAD | DTB | MEDTRONIC PUERTO RICO, INC. | 4568 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 75 YR | Other | KDR901 IMPLANTABLE PULSE GENERATOR| 5076 IMPLANTABLE PACING LEAD |