FDA Adverse Event Malfunction Summary report: N

CAPSUREFIX NOVUS

MDR report key: 1862780 · Received October 9, 2010

Report

Report Number
2649622-2010-10408
Event Type
Malfunction
Date Received
October 9, 2010
Date of Event
July 16, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039/S009
Removal / Correction Number
ASKU
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT OVERSENSING WAS BEING SEEN ON BOTH THE A AND V LEADS. IT WAS ALSO REPORTED THAT THERE WAS A SMALL DECREASE IN IMPEDANCE AND INCREASE IN CAPTURE THRESHOLD ON THE V LEAD. BOTH THE A AND V LEADS WERE CAPPED LESS THAN A WEEK LATER AND NEW LEADS WERE IMPLANTED. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 5076 ASKU

Patients

Seq Age Sex Outcome Treatment
1 65 YR Required Intervention 5071 IMPLANTABLE PACING LEAD| E2DR01AA IMPLANTABLE PACEMAKER