FDA Adverse Event
Injury
Summary report: N
CAPSURE SP
MDR report key: 1862757
·
Received October 8, 2010
Report
- Report Number
- 2649622-2010-10291
- Event Type
- Injury
- Date Received
- October 8, 2010
- Date of Event
- July 15, 2010
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- DTB
- PMA / PMN Number
- P830061/S12
- Removal / Correction Number
- ASKU
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE ATRIAL LEAD WAS PERMANENTLY PROGRAMMED TO UNIPOLAR DUE TO LOW BIPOLAR IMPEDANCE, PRESUMABLY DUE TO INSULATION BREAKDOWN. THE REPORTED BIPOLAR IMPEDANCE WAS 284 OHMS, WHILE THE UNIPOLAR IMPEDANCE WAS 324 OHMS. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CAPSURE SP | IMPLANTABLE PACING LEAD | DTB | MEDTRONIC PUERTO RICO, INC. | 4524 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 76 YR | Other | (B)(4) IMPLANTABLE PULSE GENERATOR| 5076 IMPLANTABLE PACING LEAD |