FDA Adverse Event Injury Summary report: N

CAPSURE SP

MDR report key: 1862757 · Received October 8, 2010

Report

Report Number
2649622-2010-10291
Event Type
Injury
Date Received
October 8, 2010
Date of Event
July 15, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P830061/S12
Removal / Correction Number
ASKU
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE ATRIAL LEAD WAS PERMANENTLY PROGRAMMED TO UNIPOLAR DUE TO LOW BIPOLAR IMPEDANCE, PRESUMABLY DUE TO INSULATION BREAKDOWN. THE REPORTED BIPOLAR IMPEDANCE WAS 284 OHMS, WHILE THE UNIPOLAR IMPEDANCE WAS 324 OHMS. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSURE SP IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 4524 ASKU

Patients

Seq Age Sex Outcome Treatment
1 76 YR Other (B)(4) IMPLANTABLE PULSE GENERATOR| 5076 IMPLANTABLE PACING LEAD