FDA Adverse Event Malfunction Summary report: N

COLLEAGUE CXE VOLUMETRICINFUSION PUMP

MDR report key: 1862741 · Received October 8, 2010

Report

Report Number
6000001-2010-03918
Event Type
Malfunction
Date Received
October 8, 2010
Date of Event
July 14, 2010
Report Date
July 16, 2010
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K063696
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CORRECTION: THE REPORTED CONDITION OF FAILURE CODE 500:320:654:0000 WAS CONFIRMED BUT NOT DUPLICATED. THE ROOT CAUSE HAS NOT BEEN DETERMINED AT THIS TIME. NO REPAIRS HAVE BEEN PERFORMED AT THIS TIME SINCE THIS IS A BAXTER OWNED DEVICE. A FOLLOW-UP REPORT WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE. (B)(4).

Additional Manufacturer Narrative · 1

(B)(4).THE DEVICE WAS RETURNED TO BAXTER AND IS CURRENTLY IN THE PROCESS OF BEING EVALUATED. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL DETAILS BECOME AVAILABLE.

Additional Manufacturer Narrative · 1

(B)(4). ADDITIONAL INFORMATION: THE ROOT CAUSE INVESTIGATION FOR THE REPORTED PROBLEM IS IN PROGRESS THROUGH (B)(4).

Description of Event or Problem · 1

THE FACILITY REPRESENTATIVE REPORTED A COLLEAGUE INFUSION PUMP WITH FAILURE CODE 500:320:654:0000. IT IS UNKNOWN WHEN THIS EVENT OCCURRED. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION NECESSARY. THIS DEVICE UTILIZES USER INTERFACE MODULE MASTER SOFTWARE VERSION 6.13.90 CATEGORIZED AS REMEDIATED. DURING BAXTER'S REVIEW OF THE EVENT HISTORY, IT WAS DETERMINED THAT THE REPORTED CONDITION OCCURRED DURING DELIVERY CAUSING AN INTERRUPTION OF DELIVERY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE CXE VOLUMETRICINFUSION PUMP PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1