COLLEAGUE SINGLE CHANNEL VOLUMETRIC PUMP CE FRENCH
Report
- Report Number
- 6000001-2010-03875
- Event Type
- Malfunction
- Date Received
- October 8, 2010
- Date of Event
- July 21, 2010
- Report Date
- September 15, 2010
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- Removal / Correction Number
- 6000001-12/1/08-002-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
(B)(4). A FOLLOW-UP MEDWATCH REPORT WILL BE SUBMITTED WHEN THE EVALUATION RESULTS OR IF ADDITIONAL INFORMATION BECOMES AVAILABLE. THIS DEVICE IS MANUFACTURED FOR DISTRIBUTION OUTSIDE OF THE UNITED STATES (US); THEREFORE, IT DOES NOT CONTAIN A US 510K NUMBER. HOWEVER, THIS MDR IS BEING SUBMITTED BECAUSE IT IS THE SAME AS OR SIMILAR TO A PRODUCT DISTRIBUTED WITHIN THE US.
(B)(4). A TREND REVIEW HAS BEEN PERFORMED AND THERE HAVE BEEN SIMILAR REPORTS MADE TO BAXTER. THE ROOT CAUSE WILL CONTINUE TO BE INVESTIGATED IN CAPA (CORRECTIVE AND PREVENTIVE ACTION) INVESTIGATION (B)(4).
(B)(4). DEVICE EVALUATION: THE REPORTED CONDITION WAS CONFIRMED. THE ASSIGNABLE CAUSE WAS DEPLETED MAIN BATTERIES. THE MAIN BATTERIES AND HARNESS HAVE BEEN REPLACED. THE USER INTERFACE MODULE MASTER SOFTWARE VERSION IS 5.06.00, CATEGORIZED AS UNREMEDIATED.
DURING A REVIEW OF THE EVENT HISTORY FOR ANOTHER REPORT, IT WAS DISCOVERED THAT A BATTERY DEPLETED SET ALARM OCCURRED DURING INFUSION WHICH CAUSED AN INTERRUPTION DURING DELIVERY. THERE WAS NO ADVERSE EVENT, PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS REPORT. THERE IS NO FURTHER COMPLAINT INFORMATION AVAILABLE.THE USER INTERFACE MODULE MASTER SOFTWARE VERSION FOR THIS DEVICE IS CURRENTLY UNKNOWN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE SINGLE CHANNEL VOLUMETRIC PUMP CE FRENCH | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |