FDA Adverse Event Malfunction Summary report: N

COLLEAGUE VOLUMETRIC INFUSOR PUMP-SINGLE CHANNEL

MDR report key: 1862596 · Received October 8, 2010

Report

Report Number
6000001-2010-03873
Event Type
Malfunction
Date Received
October 8, 2010
Date of Event
June 17, 2010
Report Date
September 15, 2010
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K063696
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE REPORTED CONDITION HAS BEEN CONFIRMED BUT NOT DUPLICATED. THE ASSIGNABLE CAUSE IS FAULTY PHM (PUMPHEAD MODULE) ASSEMBLY. THE PHM ASSEMBLY HAS BEEN REPLACED.

Additional Manufacturer Narrative · 1

(B)(4). A FOLLOW-UP MEDWATCH REPORT WILL BE SUBMITTED WHEN THE EVALUATION RESULTS OR ADDITIONAL INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative · 1

(B)(4). THIS DEVICE IS A REMEDIATED COLLEAGUE PUMP WITH A USER INTERFACE MODULE SOFTWARE VERSION OF 5.09.90. A SERVICE HISTORY REVIEW WAS PERFORMED REVEALING THE REPORTED CONDITION IS NOT RELATED TO ANY PREVIOUS CUSTOMER SERVICE REQUEST ON THIS PUMP. HOWEVER, DURING PREVIOUS SERVICE ON 05-23-2007, 10-27-2005, 08-04-2005 AND 12-02-2003 THE PUMPHEAD MODULE WAS REPLACED. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH MDQ-CAPA-(B)(4).

Description of Event or Problem · 1

DURING A REVIEW OF THE EVENT HISTORY FOR ANOTHER REPORT, IT WAS DISCOVERED THAT FAILURE CODE 810:03 OCCURRED TWICE ON (B)(6) 2010 DURING AN INFUSION WHICH CAUSED AN INTERRUPTION DURING DELIVERY. THERE WAS NO ADVERSE EVENT, PATIENT INJURY, OR MEDICAL INTERVENTION ASSOCIATED WITH THIS REPORT. THERE IS NO FURTHER COMPLAINT INFORMATION AVAILABLE. THE USER INTERFACE MODULE MASTER SOFTWARE VERSION IS 5.09.90, CATEGORIZED AS REMEDIATED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE VOLUMETRIC INFUSOR PUMP-SINGLE CHANNEL PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1