FDA Adverse Event Injury Summary report: N

CONCERTO CRT-D DR

MDR report key: 1862419 · Received October 8, 2010

Report

Report Number
6000144-2010-04673
Event Type
Injury
Date Received
October 8, 2010
Date of Event
June 11, 2010
Manufacturer
MEDTRONIC MED REL, INC.
Product Code
NIK
PMA / PMN Number
P010031/S031
Removal / Correction Number
Z-0126-2010
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BASED SOLELY ON DEVICE RETURN AND ANALYSIS. NO INFORMATION TO SUGGEST A DEVICE-RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE DEVICE IS PART OF THE ADVISORY FOR THIS MODEL. EVALUATION SUMMARY: (B)(4): ANALYSIS OF THE DEVICE FOUND A HIGH CURRENT DRAIN CONDITION. ELECTRICAL ANALYSIS REVEALED THE CAUSE OF THE HIGH CURRENT DRAIN WAS CURRENT LEAKAGE IN THE BATTERY FILTER CAPACITORS.

Description of Event or Problem · 1

THE DEVICE WAS RETURNED TO THE MANUFACTURER AFTER BEING EXPLANTED DUE TO NORMAL BATTERY DEPLETION. THE DEVICE SUBSEQUENTLY TESTED OUT OF SPECIFICATION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CONCERTO CRT-D DR IMPLANTABLE PACEMAKER/CARDIO/DEFIB NIK MEDTRONIC MED REL, INC. C154DWK ASKU

Patients

Seq Age Sex Outcome Treatment
1 51 YR Required Intervention 5076 IMPLANTABLE PACING LEAD| 4194 IMPLANTABLE PACING LEAD| 6949 IMPLANTABLE TACHY LEAD