FDA Adverse Event Malfunction Summary report: N

BINAXNOW COVID-19 AG SELF TEST 2CT

MDR report key: 18624177 · Received February 1, 2024

Report

Report Number
1221359-2024-00132
Event Type
Malfunction
Date Received
February 1, 2024
Date of Event
January 15, 2024
Report Date
February 26, 2024
Manufacturer
ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
Product Code
QKP
UDI-DI
00811877011408
PMA / PMN Number
EUA210264
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

D4-UDI:(B)(4) TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT 224989 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 195-160/ LOT: 224989, TEST BASE PART NUMBER 195-430H/ LOT: 220720. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 224989 SHOWED THAT THE COMPLAINT RATE IS (B)(4). ABBOTT DIAGNOSTICS SCARBOROUGH, INC. WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE. H3 OTHER TEXT : SINGLE USE; DEVICE DISCARDED

Additional Manufacturer Narrative · 0

D4-UDI:(B)(4). THE REMAINDER OF THE INVESTIGATION REMAINS IN PROGRESS. A SUPPLEMENTAL REPORT WILL BE PROVIDED AFTER COMPLETION. H3 OTHER TEXT : SINGLE USE; DEVICE DISCARDED.

Description of Event or Problem · 0

THE CONSUMER REPORTED A FALSE POSITIVE RESULT WITH THE BINAXNOW COVID-19 ANTIGEN SELF-TEST PERFORMED ON (B)(6)2024 ON AN UNKNOWN SAMPLE TYPE. ADDITIONAL TESTING WAS PERFORMED ON THE SAME DAY VIA AN UNKNOWN BRAND OF RAPID ANTIGEN TEST WHICH GENERATED A NEGATIVE RESULT. THE CONSUMER INDICATED THEY WERE ASYMPTOMATIC. NO ADDITIONAL INFORMATION, INCLUDING TREATMENT AND OUTCOME, WAS PROVIDED.

Description of Event or Problem · 0

THE CONSUMER REPORTED A FALSE POSITIVE RESULT WITH THE BINAXNOW COVID-19 ANTIGEN SELF-TEST PERFORMED ON (B)(6) 2024 ON AN UNKNOWN SAMPLE TYPE. ADDITIONAL TESTING WAS PERFORMED ON THE SAME DAY VIA AN UNKNOWN BRAND OF RAPID ANTIGEN TEST WHICH GENERATED A NEGATIVE RESULT. THE CONSUMER INDICATED THEY WERE ASYMPTOMATIC. NO ADDITIONAL INFORMATION, INCLUDING TREATMENT AND OUTCOME, WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1073169 BINAXNOW COVID-19 AG SELF TEST 2CT CORONAVIRUS ANTIGEN DETECTION TEST SYSTEM. QKP ABBOTT DIAGNOSTICS SCARBOROUGH, INC. 224989 00811877011408

Patients

Seq Age Sex Outcome Treatment
1 NA Male