FDA Adverse Event Malfunction Summary report: N

X-FLOW PROSTATECTOMY CATHETER

MDR report key: 18623610 · Received February 1, 2024

Report

Report Number
9610711-2024-00026
Event Type
Malfunction
Date Received
February 1, 2024
Date of Event
October 24, 2023
Report Date
June 4, 2024
Manufacturer
COLOPLAST A/S
Product Code
EZL
UDI-DI
03600040144572
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

AFTER RECEIVING THIS COMPLAINT, WE SEARCHED FOR OTHER COMPLAINTS AND WE FOUND NONE REGARDING THE LOT NUMBER 9132281 WITH THE SAME DEFECT. THIS PRODUCT WAS MADE BY OUR SUBCONTRACTOR WHICH WAS INFORMED ABOUT THIS ISSUE. IN FEBRUARY WE RECEIVED DOCUMENTATION OF INVESTIGATION FROM OUR SUBCONTRACTOR, THE PROBABLE ROOT CAUSE OF THIS ISSUE WAS THE RAW MATERIAL OF THE BALLOON. WE RECEIVED THE INCRIMINATED SAMPLE. AFTER DECONTAMINATION, WE NOTED THAT THE BALLOON WAS BURST WITHOUT MISSING PART. CHECKING THE QUALITY DATABASES REVEALED ONE NON-CONFORMITY IN RELATION TO THE DESCRIBED DEFECT AND A CORRECTIVE AND PREVENTIVE ACTION: - NC (B)(4): "PB BALLONS (BULLES + AGGLUTINÉS)" OPENED IN 16 JANUARY 2023. FOR THIS NC, THE USED OF CLUMPED BALLOONS SHOULD BE RESPONSIBLE OF SOME WEAKNESS ON THE BALLOON AND BEING RESPONSIBLE OF BURSTING. - MONITORING CAPA-000152: "BALLOON ISSUES ON FOLYSIL AND SILICONE PROSTATIC CATHETERS". THE TRENDING FOR BALLOON BURSTING IS SPECIFICALLY MONITORED. A SIMILAR CASE STUDY WAS PERFORMED BASED ON SAME ITEM NUMBER AB63 SAME DEFECT: BALLOON BURST OVER LAST FOUR YEAR: 53 SIMILAR CASE WERE FOUND.

Description of Event or Problem · 0

ACCORDING TO THE AVAILABLE INFORMATION THE BALLOON BURST/RUPTURED. NO SERIOUS CONSEQUENCES FOR THE PATIENT.

Description of Event or Problem · 0

ACCORDING TO THE AVAILABLE INFORMATION THE BALLOON BURST/RUPTURED. NO SERIOUS CONSEQUENCES FOR THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1109832 X-FLOW PROSTATECTOMY CATHETER INDWELLING URETHRAL DRAINAGE BALLOON CATHETER, NON-ANTIMICROBIAL EZL COLOPLAST A/S 9132281_AB63241002 03600040144572

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Required Intervention