FDA Adverse Event Injury Summary report: N

DISCOVERY II

MDR report key: 1862234 · Received October 8, 2010

Report

Report Number
2124215-2010-18898
Event Type
Injury
Date Received
October 8, 2010
Date of Event
July 28, 2010
Report Date
July 28, 2010
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
NVZ
PMA / PMN Number
D970003
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT IN OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH DEVICE EVALUATION WAS PERFORMED. NORMAL PACING AND SENSING WAS CONFIRMED. NO NOISE OR PAUSES IN PACING WERE OBSERVED DURING ADDITIONAL TESTING WHILE THE DEVICE WAS MANIPULATED. FINAL DEVICE EVALUATION WAS UNABLE TO CONFIRM THE REPORTED CLINICAL OBSERVATIONS. NO OTHER ADVERSE EVENTS HAVE BEEN REPORTED. THIS ISSUE WILL BE RE-EVALUATED IF ADDITIONAL INFORMATION IS RECEIVED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC CRM RECEIVED INFORMATION THAT THIS PATIENT PRESENTED WITH A STROKE AND A MYOCARDIAL INFARCTION (MI). SYSTEM EVALUATION NOTED NOISE ON THE RIGHT VENTRICULAR CHANNEL WHICH WAS OVERSENSED AND RESULTED IN PACING INHIBITION AND A FOUR SECOND PAUSE IN PACING FOR THIS PACEMAKER DEPENDENT PATIENT. A BOSTON SCIENTIFIC CRM TECHNICAL SERVICES CONSULTANT REVIEWED THE SYSTEM EGMS AND SUSPECTS SOME ELECTROMAGNETIC INTERFERENCE (EMI). THE CALLER REPROGRAMMED THE RIGHT VENTRICULAR SENSITIVITY TO 10 MV AND HAS WATCHED THE PATIENT TO SEE IF ANY FURTHER INHIBITION OCCURRED; WHICH HAS NOT TO THIS POINT. THE CALLER WILL DISCUSS THE CLINICAL OBSERVATIONS WITH THIS PATIENT'S PHYSICIAN. ONE MONTH LATER, THIS PACEMAKER AND RIGHT VENTRICULAR LEAD WERE REMOVED FROM SERVICE AND REPLACED WITHOUT FURTHER INCIDENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DISCOVERY II IMPLANTABLE PULSE GENERATOR NVZ GUIDANT CRM CLONMEL IRELAND 1284

Patients

Seq Age Sex Outcome Treatment
1 83 YR 4087| 1135| 1284| 0523| 4116| 4086