FDA Adverse Event
Injury
Summary report: N
FINELINE II
MDR report key: 1862171
·
Received October 8, 2010
Report
- Report Number
- 2124215-2010-16390
- Event Type
- Injury
- Date Received
- October 8, 2010
- Date of Event
- July 29, 2010
- Report Date
- July 29, 2010
- Manufacturer
- CPI - DEL CARIBE
- Product Code
- NVN
- PMA / PMN Number
- P960004
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NE
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
Additional Manufacturer Narrative · 1
THE AVAILABLE INFORMATION SUGGESTS THIS PRODUCT REMAINS IN SERVICE. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS EVENT WILL BE UPDATED.
Description of Event or Problem · 1
BOSTON SCIENTIFIC CRM RECEIVED INFORMATION THAT THIS RIGHT ATRIAL LEAD AND THE RIGHT VENTRICULAR RATE/SENSE LEAD WERE REVERSED IN THE DEVICE HEADER. THE POCKET WAS REOPENED AND THE LEADS WERE INSERTED INTO THE CORRECT PORTS. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FINELINE II | IMPLANTABLE LEAD | NVN | CPI - DEL CARIBE | 4480 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 66 YR | Required Intervention | 4480| H177| E110| 0158 |