FDA Adverse Event Malfunction Summary report: N

RECIPROC BLUE FILES, 6X, STERILE

MDR report key: 18621423 · Received February 1, 2024

Report

Report Number
9611053-2024-00013
Event Type
Malfunction
Date Received
February 1, 2024
Report Date
February 26, 2024
Manufacturer
VDW GMBH
Product Code
EKS
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
DENTIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

SINCE THIS EVENT RESULTED IN MEDICAL/SURGICAL INTERVENTION TO PRECLUDE PERMANENT DAMAGE TO A BODY STRUCTURE OR PERMANENT IMPAIRMENT OF A BODY FUNCTION, IT MUST BE PRESUMED THAT THE MALFUNCTION WOULD BE LIKELY TO CAUSE OR CONTRIBUTE TO A SERIOUS INJURY SHOULD IT RECUR. AS SUCH, THIS EVENT MEETS THE DEFINITION OF A REPORTABLE EVENT PER 21 CFR PART 803. THE DEVICE IS AVAILABLE FOR EVALUATION, THOUGH RESULTS ARE NOT AVAILABLE AS OF THIS REPORT. EVALUATION RESULTS WILL BE SUBMITTED AS THEY BECOME AVAILABLE.

Additional Manufacturer Narrative · 0

SUMMARY: INVOLVED PRODUCTS WHICH BROKE DURING USE WERE NOT RETURNED AND CANNOT BE ANALYZED. NOTHING UNUSUAL TO REPORT WAS FOUND DURING DHRS REVIEW (BATCHES #1760628, #1760931, #1763768, #1760627, #1761739, #1764586, #1765158, #1764627, #1765624, #1764600 AND #1763562). THE TWO UNUSED RECIPROC BLUE FILES R25 8/100 25MM 025 WHICH WERE RETURNED WERE EVALUATED AND WERE FOUND IN COMPLIANCE WITH SPECIFICATIONS. ROOT CAUSES ARE NOT IDENTIFIED. WE WILL TRACK THIS KIND OF EVENT AND MONITOR THE TREND. FOR INFORMATION, WE REMIND THE PRACTITIONER HAS TO MAKE SURE THAT A STRAIGHT-LINE ACCESS IS CREATED PRIOR USING THE RECIPROC BLUE FILES (AS MENTIONED IN THE DFU).

Description of Event or Problem · 0

IN THIS EVENT IT IS REPORTED THAT RECIPROC BLUE FILES, 6X, STERILE BROKE DURING USE. THE TOOTH WITH THIS BROKEN PART WAS EXTRACTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
374454 RECIPROC BLUE FILES, 6X, STERILE FILE, PULP CANAL, ENDODONTIC EKS VDW GMBH 390052

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown