FDA Adverse Event Malfunction Summary report: N

CONTAK RENEWAL

MDR report key: 1862141 · Received October 8, 2010

Report

Report Number
2124215-2010-17203
Event Type
Malfunction
Date Received
October 8, 2010
Date of Event
July 29, 2010
Report Date
July 29, 2010
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
NIK
Removal / Correction Number
Z-1192-06 THRU Z-1194-06
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

A REVIEW OF DEVICE MEMORY REVEALED THAT THE DEVICE DECLARED END OF LIFE (EOL) AFTER EXPERIENCING ONE CHARGE TIME GREATER THAN 30 SECONDS. THE OBSERVED RATE OF BATTERY USAGE WAS COMPARED TO THE EXPECTED RATE OF BATTERY USAGE; THE RESULTS INDICATED THAT THE MONITORING VOLTAGE WAS NORMAL. ALTHOUGH THE BATTERY ITSELF HAD NOT DEPLETED PREMATURELY, EOL WAS TRIGGERED EARLIER THAN EXPECTED BY EXTENDED CHARGE TIMES DUE TO A HIGHER-THAN-TYPICAL BUILD-UP OF INTERNAL BATTERY IMPEDANCE. THIS ISSUE IS DISCUSSED IN THE (B)(4) 2010 PRODUCT PERFORMANCE REPORT.

Description of Event or Problem · 1

BOSTON SCIENTIFIC CRM RECEIVED INFORMATION THAT THIS DEVICE WAS EXPLANTED AND RETURNED FOR DISPOSAL. THERE ARE NO PRODUCT PERFORMANCE ALLEGATIONS OR ADVERSE PATIENT EFFECTS REGARDING THIS DEVICE. ROUTINE INITIAL RETURNED DEVICE TESTING INDICATED THAT DETAILED ANALYSIS WAS REQUIRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CONTAK RENEWAL IMPLANTABLE CHF GENERATOR NIK GUIDANT CRM CLONMEL IRELAND H170

Patients

Seq Age Sex Outcome Treatment
1 62 YR H170| 5076| 0184| E110