FDA Adverse Event Malfunction Summary report: N

K133801

MDR report key: 18619408 · Received January 31, 2024

Report

Report Number
1314492-2024-00224
Event Type
Malfunction
Date Received
January 31, 2024
Report Date
March 8, 2024
Manufacturer
BAXTER HEALTHCARE CORPORATION
Product Code
FRN
UDI-DI
00085412498683
PMA / PMN Number
K133801
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

E1: INITIAL REPORTER PHONE NO. - (B)(6). SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION: D9, H3, H6, H10. H10: THE DEVICE WAS RECEIVED FOR EVALUATION. DURING FUNCTIONAL TESTING, THE REPORTED PROBLEM OF "DOOR FULLY LATCHED" WAS REPRODUCED. A SERVICE HISTORY REVIEW REVEALED NO INDICATION THAT THE PARTS REPLACED DURING SERVICING CAUSED OR CONTRIBUTED TO THE REPORTED EVENT. THE REPORTED CONDITION WAS VERIFIED. THE CAUSE OF THE CONDITION WAS DETERMINED TO BE THE LOWER AUXILIARY FAILURE. THE LOWER AUXILIARY REQUIRES REPLACEMENT TO ADDRESS THIS ISSUE. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 0

IT WAS REPORTED THAT A SPECTRUM PUMP ALARMED " DOOR NOT CLOSED". THIS OCCURRED DURING TESTING IN THE BIOMEDICAL DEPARTMENT. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
414705 K133801 PUMP, INFUSION FRN BAXTER HEALTHCARE CORPORATION NA NA 00085412498683

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown