CONCERTO CRT-D DR
Report
- Report Number
- 2647346-2010-00585
- Event Type
- Malfunction
- Date Received
- October 8, 2010
- Date of Event
- June 16, 2010
- Manufacturer
- MEDTRONIC MED REL, INC.
- Product Code
- NIK
- PMA / PMN Number
- P010031/S031
- Removal / Correction Number
- ASKU
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) ACTUAL LONGEVITY IS < 80% OF 99.9% LONGEVITY LIMIT. THE DEVICE WAS FULLY FUNCTIONAL WITH NO HIGH CURRENT DRAIN OR EVIDENCE OF BATTERY PROBLEMS. WITHOUT THE HISTORY OF THE PROGRAMMED SETTINGS THROUGHOUT ITS SERVICE LIFE, THERE IS NO WAY TO DETERMINE WHY THE LONGEVITY DID NOT MATCH THE PREDICTED MODEL.
IT WAS REPORTED THAT BOTH LEADS HAD INCREASED THRESHOLDS (RIGHT VENTRICULAR AND LEFT VENTRICULAR). BOTH LEADS ARE STILL IN USE. NO FURTHER PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE DEVICE WAS RETURNED, ANALYZED AND SUBSEQUENTLY TESTED OUT OF SPECIFICATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CONCERTO CRT-D DR | IMPLANTABLE PACEMAKER/CARDIO/DEFIB | NIK | MEDTRONIC MED REL, INC. | C154DWK | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 59 YR | Required Intervention |