FDA Adverse Event Malfunction Summary report: N

BD LUER-LOK

MDR report key: 18618279 · Received January 31, 2024

Report

Report Number
1213809-2024-00048
Event Type
Malfunction
Date Received
January 31, 2024
Date of Event
January 10, 2024
Report Date
July 17, 2024
Manufacturer
BECTON DICKINSON MEDICAL SYSTEMS
Product Code
FMF
UDI-DI
30382903096054
PMA / PMN Number
K110771
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ONE PHOTO OF 10ML TRAY PACKAGES LOT 3152766 WAS RECEIVED AND EVALUATED. THE IMAGE SHOWS THREE SEALED TRAY PACKAGES WITH ONLY THE LOT, UDI, AND EXPIRATION DATE INFORMATION ON ONE OF THE TRAY PACKAGES AND MISSING FROM THE TWO OTHERS. THE CONDITION OBSERVED IS NON-CONFORMING PER PRODUCT SPECIFICATION. POTENTIAL ROOT CAUSE FOR THE LABEL CONTENT MISSING DEFECT IS ASSOCIATED WITH THE PACKAGING PROCESS. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED LOT NUMBER 3152766 SHOWING NO REJECTED INSPECTIONS OR QUALITY ISSUES DURING THE PRODUCTION OF THE PROVIDED LOT NUMBER THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT.

Additional Manufacturer Narrative · 0

H.3. IF A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

MATERIALS#: 309605, BATCH#: UNKNOWN. IT WAS REPORTED BY THE CUSTOMER THAT ONE OF THE BOXES TWO PACKS WITH NO LOT NUMBERS OR EXPIRY ON THEM. VERBATIM: THIS TIME I'M REACHING OUT TO YOU FOR A DIFFERENT MATTER REGARDING THE 10ML SYRINGE TRAY WITH PRODUCT CODE 309605. WE ORDER THIS PRODUCT FROM CARDINAL HEALTH IN LARGE QUANTITIES. IN THE MOST RECENT ORDER, WE FOUND IN ONE OF THE BOXES TWO PACKS WITH NO LOT NUMBERS OR EXPIRY ON THEM.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD LUER-LOK LABEL CONTENT WAS MISSING. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "THIS TIME I'M REACHING OUT TO YOU FOR A DIFFERENT MATTER REGARDING THE 10ML SYRINGE TRAY WITH PRODUCT CODE 309605 . WE ORDER THIS PRODUCT FROM CARDINAL HEALTH IN LARGE QUANTITIES. IN THE MOST RECENT ORDER, WE FOUND IN ONE OF THE BOXES TWO PACKS WITH NO LOT NUMBERS OR EXPIRY ON THEM."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1261138 BD LUER-LOK PISTON SYRINGE FMF BECTON DICKINSON MEDICAL SYSTEMS 3152766 30382903096054

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown