FDA Adverse Event Injury Summary report: N

SIGMA 300 S

MDR report key: 1861796 · Received October 8, 2010

Report

Report Number
6000024-2010-00056
Event Type
Injury
Date Received
October 8, 2010
Date of Event
June 24, 2010
Manufacturer
MEDTRONIC OF CANADA LTD.
Product Code
DXY
PMA / PMN Number
P980035/S2
Removal / Correction Number
Z-1510/1519-2009
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY: (B)(4) TESTING REVEALED NO OUTPUT AND NO TELEMETRY. THE NO OUTPUT AND NO TELEMETRY CONDITIONS WERE THE RESULT OF LIFTED HYBRID BOND WIRES. THE DEVICE IS PART OF THE ADVISORY FOR THIS MODEL.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE HAD NO PULSE SIGNAL WHEN PROGRAMMING. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SIGMA 300 S IMPLANTABLE PULSE GENERATOR DXY MEDTRONIC OF CANADA LTD. SS303 ASKU

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention