FDA Adverse Event
Malfunction
Summary report: N
SPRINT QUATTRO SECURE
MDR report key: 1861790
·
Received October 8, 2010
Report
- Report Number
- 2649622-2010-10094
- Event Type
- Malfunction
- Date Received
- October 8, 2010
- Date of Event
- July 9, 2010
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- LWS
- PMA / PMN Number
- P920015/S24
- Removal / Correction Number
- ASKU
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE LEAD WAS RETURNED WITH THE HELIX HUNG UP ON THE GUIDE TOOTH. FULL LEAD RETURNED AND ANALYZED. BLOOD IN/ON DISTAL CONDUCTOR (NOT OBSTRUCTED) AND IN/ON HELIX MECHANISM (SLEEVE HEAD). HELIX DISENGAGED FROM HELICAL CHANNEL.
Description of Event or Problem · 1
IT WAS REPORTED THAT THAT DURING AN IMPLANT OF A RIGHT VENTRICULAR LEAD, THE "SCREW" MECHANISM OF THE LEAD WAS BROKEN. THE LEAD WAS ATTEMPTED AND NOT USED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPRINT QUATTRO SECURE | IMPLANTABLE TACHY LEAD | LWS | MEDTRONIC PUERTO RICO, INC. | 6947 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 53 YR | Required Intervention |