SPRINT QUATTRO SECURE
Report
- Report Number
- 2649622-2010-10115
- Event Type
- Malfunction
- Date Received
- October 8, 2010
- Date of Event
- July 8, 2010
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- LWS
- PMA / PMN Number
- P920015/S24
- Removal / Correction Number
- ASKU
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4): THE ACTUAL DEVICE WAS NOT RECEIVED FOR EVALUATION. WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA. HIGHEST WEEKLY TREND FOR RIGHT VENTRICULAR PACE IMPEDANCE IS 1296 FROM THE WEEK OF (B)(4) 2010. A RECENT DAILY VALUE OF 1680 OHMS WAS RECORDED ON (B)(4) 2010. THE WEEKLY VALUE CORRESPONDING TO THAT DAILY VALUE HAS NOT YET BEEN ADDED TO THE WEEKLY TREND. NO ANOMALIES WERE FOUND.
ASKU
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT THE PATIENT CAME INTO THE OFFICE DUE TO PATIENT ALERT TONES FOR OUT OF RANGE IMPEDANCE. THE CALLER STATED THAT THE PACING LEAD IMPEDANCE REPORTED AN OBSERVATION TEXT AS 1680 OHMS ALTHOUGH THE LEAD IMPEDANCE TRENDS ONLY SHOWED HIGHEST AS 1296 OHMS. IT WAS FURTHER REPORTED THAT CROSS TALK SEEMED TO BE OCCURRING WITH THE LEAD. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
IT WAS REPORTED THAT THE PATIENT CAME INTO THE OFFICE DUE TO PATIENT ALERT TONES FOR OUT OF RANGE IMPEDANCE. THE CALLER STATED THAT THE PACING LEAD IMPEDANCE REPORTED AN OBSERVATION TEXT AS 1680 OHMS ALTHOUGH THE LEAD IMPEDANCE TRENDS ONLY SHOWED HIGHEST AS 1296 OHMS. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPRINT QUATTRO SECURE | ASKU | LWS | MEDTRONIC PUERTO RICO, INC. | 6947 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 64 YR | Other | 5076 IMPLANTABLE PACING LEAD| (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 4968 IMPLANTABLE PACING LEAD| 4968 IMPLANTABLE PACING LEAD| (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 5076 IMPLANTABLE PACING LEAD| 4968 IMPLANTABLE PACING LEAD| 5076 IMPLANTABLE PACING LEAD |