FDA Adverse Event Malfunction Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 1861757 · Received October 8, 2010

Report

Report Number
2649622-2010-10115
Event Type
Malfunction
Date Received
October 8, 2010
Date of Event
July 8, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S24
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4): THE ACTUAL DEVICE WAS NOT RECEIVED FOR EVALUATION. WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA. HIGHEST WEEKLY TREND FOR RIGHT VENTRICULAR PACE IMPEDANCE IS 1296 FROM THE WEEK OF (B)(4) 2010. A RECENT DAILY VALUE OF 1680 OHMS WAS RECORDED ON (B)(4) 2010. THE WEEKLY VALUE CORRESPONDING TO THAT DAILY VALUE HAS NOT YET BEEN ADDED TO THE WEEKLY TREND. NO ANOMALIES WERE FOUND.

Additional Manufacturer Narrative · 1

ASKU

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT CAME INTO THE OFFICE DUE TO PATIENT ALERT TONES FOR OUT OF RANGE IMPEDANCE. THE CALLER STATED THAT THE PACING LEAD IMPEDANCE REPORTED AN OBSERVATION TEXT AS 1680 OHMS ALTHOUGH THE LEAD IMPEDANCE TRENDS ONLY SHOWED HIGHEST AS 1296 OHMS. IT WAS FURTHER REPORTED THAT CROSS TALK SEEMED TO BE OCCURRING WITH THE LEAD. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT CAME INTO THE OFFICE DUE TO PATIENT ALERT TONES FOR OUT OF RANGE IMPEDANCE. THE CALLER STATED THAT THE PACING LEAD IMPEDANCE REPORTED AN OBSERVATION TEXT AS 1680 OHMS ALTHOUGH THE LEAD IMPEDANCE TRENDS ONLY SHOWED HIGHEST AS 1296 OHMS. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE ASKU LWS MEDTRONIC PUERTO RICO, INC. 6947 ASKU

Patients

Seq Age Sex Outcome Treatment
1 64 YR Other 5076 IMPLANTABLE PACING LEAD| (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 4968 IMPLANTABLE PACING LEAD| 4968 IMPLANTABLE PACING LEAD| (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 5076 IMPLANTABLE PACING LEAD| 4968 IMPLANTABLE PACING LEAD| 5076 IMPLANTABLE PACING LEAD