FDA Adverse Event Injury Summary report: N

ATTAIN OTW

MDR report key: 1861749 · Received October 8, 2010

Report

Report Number
2649622-2010-10122
Event Type
Injury
Date Received
October 8, 2010
Date of Event
July 2, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P010015/S012
Removal / Correction Number
ASKU
Report Source
Manufacturer report
Reporter Location
DE, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY; (B)(4) NO ANOMALIES FOUND. PROXIMAL CONDUCTOR HAD BLOOD/BODY FLUID, OUTER INSULATION BREACHED CUT. THE FULL LEAD WAS RETURNED AND ANALYZED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEAD WAS REMOVED AND REPLACED DUE TO HIGH THRESHOLDS. UNABLE TO GET GOOD THRESHOLDS WITH PATIENT'S ANATOMY. A NEW LEAD WAS PLACED IN A NEW LOCATION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ATTAIN OTW IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 4194 ASKU

Patients

Seq Age Sex Outcome Treatment
1 67 YR Required Intervention 4076 IMPLANTABLE PACING LEAD| 6949 IMPLANTABLE TACHY LEAD| (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB