FDA Adverse Event
Injury
Summary report: N
ATTAIN OTW
MDR report key: 1861749
·
Received October 8, 2010
Report
- Report Number
- 2649622-2010-10122
- Event Type
- Injury
- Date Received
- October 8, 2010
- Date of Event
- July 2, 2010
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- DTB
- PMA / PMN Number
- P010015/S012
- Removal / Correction Number
- ASKU
- Report Source
- Manufacturer report
- Reporter Location
- DE, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY; (B)(4) NO ANOMALIES FOUND. PROXIMAL CONDUCTOR HAD BLOOD/BODY FLUID, OUTER INSULATION BREACHED CUT. THE FULL LEAD WAS RETURNED AND ANALYZED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE LEAD WAS REMOVED AND REPLACED DUE TO HIGH THRESHOLDS. UNABLE TO GET GOOD THRESHOLDS WITH PATIENT'S ANATOMY. A NEW LEAD WAS PLACED IN A NEW LOCATION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ATTAIN OTW | IMPLANTABLE PACING LEAD | DTB | MEDTRONIC PUERTO RICO, INC. | 4194 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 67 YR | Required Intervention | 4076 IMPLANTABLE PACING LEAD| 6949 IMPLANTABLE TACHY LEAD| (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB |