PD CYCLER 110 VOLT HOME CHOICEPRO AUTOMATED
Report
- Report Number
- 1423500-2010-04135
- Event Type
- Malfunction
- Date Received
- October 8, 2010
- Date of Event
- August 7, 2010
- Report Date
- September 16, 2010
- Manufacturer
- BAXTER HEALTHCARE - LARGO
- Product Code
- FKX
- PMA / PMN Number
- K053512
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE DEVICE HAS BEEN RETURNED TO THE BAXTER PRODUCT ANALYSIS LAB FOR EVALUATION. UPON COMPLETION OF THE EVALUATION A FOLLOW UP REPORT WILL BE SUBMITTED.
(B)(4). THE PRODUCT ANALYSIS LAB (PAL) EVALUATION AND HOMECHOICE (HC) DESIGN ENGINEERS AT (B)(4) OF THIS INCIDENT CONCUR THAT NO DEVICE FAILURE OR MALFUNCTION WAS IDENTIFIED THAT COULD HAVE CONTRIBUTED TO THE 18 IIPVS FOUND IN THE DEVICE'S EVENT LOGS. THE DEVICE MET PERFORMANCE SPECIFICATION REQUIREMENTS RELATIVE TO THE ISSUES OF IIPV. THE PRODUCT ANALYSIS LAB (PAL) EVALUATION AND HOMECHOICE (HC) DESIGN ENGINEERS AT (B)(4) OF THIS INCIDENT CONCUR THAT NO DEVICE FAILURE OR MALFUNCTION WAS IDENTIFIED THAT COULD HAVE CONTRIBUTED TO THE 18 IIPVS FOUND IN THE DEVICE'S EVENT LOGS. THE DEVICE MET PERFORMANCE SPECIFICATION REQUIREMENTS RELATIVE TO THE ISSUES OF IIPV. THE DEVICE HAS NOT BEEN PREVIOUSLY RETURNED FOR ANY ISSUES RELATED TO PEDIATRIC IIPV OR TIMEOUT DURING RITE. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4). THIS IS REPORT 12 OF 18 RELATED TO IIPV INCIDENTS FOUND IN THE DEVICE'S EVENT LOGS.
AN INCREASED INTRAPERITONEAL VOLUME (IIPV) SITUATION WAS IDENTIFIED IN THE LOG OF A RETURNED HOMECHOICE (HC) DEVICE. THE IIPV OCCURRED ON (B)(6) 2010 DURING DRAIN CYCLE 6. THE PROGRAMMED FILL VOLUME WAS 140ML AND THE TOTAL DRAIN VOLUME WAS 207ML. THIS DRAIN VOLUME MEETS BAXTER'S IIPV CRITERIA. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS REPORTED. THIS EVENT IS THE TWELFTH OF 18 EVENTS OF IIPV NOTED IN THE PATIENT LOGS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PD CYCLER 110 VOLT HOME CHOICEPRO AUTOMATED | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - LARGO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 14 MO |