FDA Adverse Event Malfunction Summary report: N

MAXIMO II DR

MDR report key: 1861663 · Received October 8, 2010

Report

Report Number
6000144-2010-04763
Event Type
Malfunction
Date Received
October 8, 2010
Date of Event
June 24, 2010
Manufacturer
MEDTRONIC MED REL, INC.
Product Code
LWS
PMA / PMN Number
P980016/S114
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE ACTUAL DEVICE WAS NOT RECEIVED FOR EVALUATION. THE PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS ANALYZED AND THE DATA REVEALED POWER ON RESET - (B)(4).

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE ACTUAL DEVICE WAS NOT RECEIVED FOR EVALUATION. THE PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS ANALYZED AND THE DATA REVEALED POWER ON RESET - POWER ON RESET PARAMETERS. WRITE TO LOCKED RAM POR RESIGTERED ON (B)(6) 2010 WITH ECC-LIA CONFLICT AT ADDRESS 69 CA. ADDITIONAL PARITY ERRORS ARE OBSERVED ON (B)(6) 2010. PATIENT ALERT LIST CONFIRMS "DEVICE CIRUIT ERROR" ON (B)(6) 2010.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE HAD AN ELECTRICAL RESET DUE TO A MEMORY ERROR. THE DEVICE IS STILL IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MAXIMO II DR ASKU LWS MEDTRONIC MED REL, INC. D284DRG ASKU

Patients

Seq Age Sex Outcome Treatment
1 78 YR Other 6947 IMPLANTABLE TACHY LEAD| 4076 IMPLANTABLE PACING LEAD| 4076 IMPLANTABLE PACING LEAD| 6947 IMPLANTABLE TACHY LEAD