MAXIMO II DR
Report
- Report Number
- 6000144-2010-04763
- Event Type
- Malfunction
- Date Received
- October 8, 2010
- Date of Event
- June 24, 2010
- Manufacturer
- MEDTRONIC MED REL, INC.
- Product Code
- LWS
- PMA / PMN Number
- P980016/S114
- Removal / Correction Number
- ASKU
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE ACTUAL DEVICE WAS NOT RECEIVED FOR EVALUATION. THE PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS ANALYZED AND THE DATA REVEALED POWER ON RESET - (B)(4).
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE ACTUAL DEVICE WAS NOT RECEIVED FOR EVALUATION. THE PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS ANALYZED AND THE DATA REVEALED POWER ON RESET - POWER ON RESET PARAMETERS. WRITE TO LOCKED RAM POR RESIGTERED ON (B)(6) 2010 WITH ECC-LIA CONFLICT AT ADDRESS 69 CA. ADDITIONAL PARITY ERRORS ARE OBSERVED ON (B)(6) 2010. PATIENT ALERT LIST CONFIRMS "DEVICE CIRUIT ERROR" ON (B)(6) 2010.
IT WAS REPORTED THAT THE DEVICE HAD AN ELECTRICAL RESET DUE TO A MEMORY ERROR. THE DEVICE IS STILL IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MAXIMO II DR | ASKU | LWS | MEDTRONIC MED REL, INC. | D284DRG | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 78 YR | Other | 6947 IMPLANTABLE TACHY LEAD| 4076 IMPLANTABLE PACING LEAD| 4076 IMPLANTABLE PACING LEAD| 6947 IMPLANTABLE TACHY LEAD |