FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 1861592 · Received October 8, 2010

Report

Report Number
2649622-2010-10024
Event Type
Injury
Date Received
October 8, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S24
Removal / Correction Number
ASKU
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT RECEIVED TWO INAPPROPRIATE SHOCKS DUE TO T-WAVE OVERSENSING. T-WAVE OVERSENSING WAS ONLY SEEN AT FASTER INTRINSIC RATES. AN INQUIRY WAS MADE TO SEE WHAT COULD BE PROGRAMMED TO PREVENT THE OVERSENSING. THE LEAD REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6947 ASKU

Patients

Seq Age Sex Outcome Treatment
1 54 YR Required Intervention (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 5076 IMPLANTABLE PACING LEAD