FDA Adverse Event Injury Summary report: N

ASKU

MDR report key: 1861561 · Received October 8, 2010

Report

Report Number
2182208-2010-00722
Event Type
Injury
Date Received
October 8, 2010
Manufacturer
MEDTRONIC, INC.
Product Code
DTB
PMA / PMN Number
ASKU
Removal / Correction Number
ASKU
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BASED SOLELY ON DEVICE RETURN AND ANALYSIS. NO INFORMATION TO SUGGEST A DEVICE-RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. WITHOUT A LOT NUMBER OR DEVICE SERIAL NUMBER, THE MANUFACTURING DATE CANNOT BE DETERMINED. EVALUATION SUMMARY: INNER INSULATION BREACHED. ALL CONDUCTORS HAD BLOOD, OUTER INSULATION KINKED/BUCKLED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

THE LEAD WAS RETURNED TO THE MANUFACTURER WITH NO INFORMATION, ANALYZED AND SUBSEQUENTLY TESTED OUT OF SPECIFICATION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ASKU IMPLANTABLE PACING LEAD DTB MEDTRONIC, INC. LEAD/MEDT ASKU

Patients

Seq Age Sex Outcome Treatment
1 Other