FDA Adverse Event
Injury
Summary report: N
VIRTUOSO DR
MDR report key: 1861541
·
Received October 8, 2010
Report
- Report Number
- 6000144-2010-04806
- Event Type
- Injury
- Date Received
- October 8, 2010
- Manufacturer
- MEDTRONIC MED REL, INC.
- Product Code
- LWS
- PMA / PMN Number
- P980016/S062
- Removal / Correction Number
- Z-0128-2010
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) ANALYSIS OF THE DEVICE FOUND A HIGH CURRENT DRAIN CONDITION. ELECTRICAL ANALYSIS REVEALED THE CAUSE OF THE HIGH CURRENT DRAIN WAS CURRENT LEAKAGE IN THE BATTERY FILTER CAPACITORS. THE DEVICE IS PART OF THE ADVISORY FOR THIS MODEL.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE DEVICE HAD REACHED ITS RECOMMENDED REPLACEMENT TIME (RRT) AND, BASED ON THE OUTPUT SETTINGS AND USAGE, THIS WAS REPORTED TO BE SOONER THAN EXPECTED. THE DEVICE WAS EXPLANTED AND REPLACED. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VIRTUOSO DR | IMPLANTABLE PACEMAKER/CARDIO/DEFIB | LWS | MEDTRONIC MED REL, INC. | D154AWG | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 77 YR | Required Intervention | 4068 IMPLANTABLE PACING LEAD| 6945 IMPLANTABLE TACHY LEAD |