FDA Adverse Event Malfunction Summary report: N

BD PRESET¿

MDR report key: 18614850 · Received January 31, 2024

Report

Report Number
9617032-2024-00105
Event Type
Malfunction
Date Received
January 31, 2024
Date of Event
January 11, 2024
Report Date
February 22, 2024
Manufacturer
BECTON, DICKINSON AND COMPANY (BD)
Product Code
JKA
UDI-DI
50382903644167
PMA / PMN Number
K022426
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H.6 INVESTIGATION SUMMARY. MATERIAL #: 364416. LOT/BATCH #: 2152474. BD HAD NOT RECEIVED SAMPLES, BUT 2 PHOTOS WERE PROVIDED FOR INVESTIGATION. THE PHOTOS WERE REVIEWED AND THE INDICATED FAILURE MODE FOR AIR BUBBLES WAS OBSERVED. ADDITIONALLY, 10 RETENTION SAMPLES FROM BD INVENTORY WERE EVALUATED BY VISUAL EXAMINATION AND FUNCTIONAL TESTING, EACH DRAWN WITH WATER, AND THE ISSUE OF AIR BUBBLES WAS NOT OBSERVED. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT HAS BEEN CONFIRMED FOR THE INDICATED FAILURE MODE AIR BUBBLES. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED.

Additional Manufacturer Narrative · 0

E.1. INITIAL REPORTER ADDR 1: (B)(6) H.6 INVESTIGATION SUMMARY: MATERIAL #: 364416 LOT/BATCH #: 2152474 BD HAD NOT RECEIVED SAMPLES OR PHOTOS FOR INVESTIGATION. THEREFORE, 10 RETENTION SAMPLES FROM BD INVENTORY WERE EVALUATED BY VISUAL EXAMINATION AND FUNCTIONAL TESTING, EACH DRAWN WITH WATER, AND NO ISSUES WERE OBSERVED RELATING TO AIR BUBBLES AS ALL SAMPLES MET SPECIFICATIONS. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT IS UNABLE TO BE CONFIRMED FOR THE INDICATED FAILURE MODE AIR BUBBLES. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS. H3 OTHER TEXT : SEE H.10

Description of Event or Problem · 0

IT WAS REPORTED THAT WHEN USING THE BD PRESET¿, AIR BUBBLES WERE PRESENT IN ONE OF THE SYRINGES. NO PATIENT IMPACT OR INJURY REPORTED.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHEN USING THE BD PRESET¿, AIR BUBBLES WERE PRESENT IN ONE OF THE SYRINGES. NO PATIENT IMPACT OR INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1245838 BD PRESET¿ BLOOD SPECIMEN COLLECTION DEVICE JKA BECTON, DICKINSON AND COMPANY (BD) 2152474 50382903644167

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown