FDA Adverse Event Injury Summary report: N

CRYSTALLINE

MDR report key: 1861452 · Received October 8, 2010

Report

Report Number
2649622-2010-10075
Event Type
Injury
Date Received
October 8, 2010
Date of Event
July 5, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039/S017
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE.

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY: (B)(4) THE ROOT CAUSE OF THE CONFIRMED EGM NOISE COULD NOT BE DETERMINED. ANALYSIS FOUND A CRACKED INTEGRATED CIRCUIT ON THE HYBRID. IT COULD NOT BE DETERMINED WHEN THE CRACK OCCURRED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE ATRIAL AND RIGHT VENTRICULAR ELECTROGRAMS SHOWED ELECTROMAGNETIC INTERFERENCE WHILE THE PACEMAKER PERFORMED APPROPRIATE SENSING WITH NO OVERSENSING. THE SURFACE ELECTROGRAM DID NOT SHOW ELECTROMAGNETIC INTERFENENCE, SO THE PHYSICIAN DID NOT BELIEVE THE SOURCE OF THE INTERFERENCE WAS FROM AN OUTSIDE SOURCE. THE DEVICE WAS EXPLANTED AND REPLACED. THE LEADS ARE STILL IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CRYSTALLINE ASKU DTB MEDTRONIC PUERTO RICO, INC. ICQ09B ASKU

Patients

Seq Age Sex Outcome Treatment
1 65 YR Hospitalization| R