FDA Adverse Event
Malfunction
Summary report: N
MONOPOLAR CURVED SCISSORS INSTRUMENT
MDR report key: 1861380
·
Received October 8, 2010
Report
- Report Number
- 2955842-2010-00427
- Event Type
- Malfunction
- Date Received
- October 8, 2010
- Date of Event
- September 7, 2010
- Report Date
- September 7, 2010
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050005
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE INSTRUMENT WAS RETURNED WITH THE DETACHED BLADE AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING OBSERVED THAT THE BLADE FRACTURED AT THE CROSS SECTION OF THE GRIP CABLE CRIMP. THE CABLE CRIMP HAS SLIPPED OUT FROM CRIMP POCKET AND THE FRACTURED PLANE OF THE BLADE IS NEARLY STRAIGHT. NO OTHER DAMAGE FOUND.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING A DA VINCI S SURGICAL PROCEDURE, A PART OF A BLADE FROM THE MONOPOLAR CURVED SCISSORS INSTRUMENT BROKE AND FELL INTO THE PATIENT. THE PIECE WAS RETRIEVED AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MONOPOLAR CURVED SCISSORS INSTRUMENT | ENDOSCOPIC ELECTROSURGICAL INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420179-09 | M10100121 371 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | DA VINCI S SYSTEM, ACCESSORIES & ESU |