FDA Adverse Event Malfunction Summary report: N

MONOPOLAR CURVED SCISSORS INSTRUMENT

MDR report key: 1861380 · Received October 8, 2010

Report

Report Number
2955842-2010-00427
Event Type
Malfunction
Date Received
October 8, 2010
Date of Event
September 7, 2010
Report Date
September 7, 2010
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050005
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED WITH THE DETACHED BLADE AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING OBSERVED THAT THE BLADE FRACTURED AT THE CROSS SECTION OF THE GRIP CABLE CRIMP. THE CABLE CRIMP HAS SLIPPED OUT FROM CRIMP POCKET AND THE FRACTURED PLANE OF THE BLADE IS NEARLY STRAIGHT. NO OTHER DAMAGE FOUND.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI S SURGICAL PROCEDURE, A PART OF A BLADE FROM THE MONOPOLAR CURVED SCISSORS INSTRUMENT BROKE AND FELL INTO THE PATIENT. THE PIECE WAS RETRIEVED AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MONOPOLAR CURVED SCISSORS INSTRUMENT ENDOSCOPIC ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420179-09 M10100121 371

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI S SYSTEM, ACCESSORIES & ESU