FDA Adverse Event Malfunction Summary report: N

HOMECHOICE CYCLER-REFURBISHED

MDR report key: 1861363 · Received October 8, 2010

Report

Report Number
1423500-2010-04158
Event Type
Malfunction
Date Received
October 8, 2010
Date of Event
September 1, 2010
Report Date
September 14, 2010
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
PMA / PMN Number
K053512
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED AND EVALUATED BY THE PRODUCT ANALYSIS LAB. THE DEVICE PASSED BOTH THE HOMECHOICE RETURN INSTRUMENT TEST / EVALUATION (RITE) ELECTRICAL TEST AND THE RITE FUNCTIONAL TEST. THE PAL EVALUATED THE DEVICE AND NO FAILURE OR MALFUNCTION WAS IDENTIFIED THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED DIFFICULTIES OR THE IIPV FOUND IN THE DEVICE LOGS. THE DEVICE WAS DETERMINED TO MEET SPECIFICATIONS. THE CAUSE OF THE INCREASED INTRA-PERITONEAL VOLUME (IIPV) IDENTIFIED IN THE DEVICE LOG WAS DETERMINED TO BE INSUFFICIENT DRAIN, FALSE EMPTY DETECT AT INITIAL DRAIN. A SERVICE HISTORY REVIEW REVEALED NO PREVIOUS SERVICE EVENTS WERE RELATED TO THE REPORTED CONDITION. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION EXPECTED BUT NOT YET COMPLETED. ANY RESULTS OF EVALUATION WILL BE PROVIDED IN A FOLLOW UP EMDR.

Description of Event or Problem · 1

DURING INITIAL ASSESSMENT OF A RETURNED HOMECHOICE (HC) DEVICE, A BAXTER TECHNICIAN IDENTIFIED AN INCREASED INTRAPERITONEAL VOLUME (IIPV) EVENT WHICH OCCURRED ON (B)(6) 2010 DURING DRAIN CYCLE 1. THE ULTRAFILTRATION (UF) VOLUME WAS 1333ML. THE PROGRAMMED FILL VOLUME WAS 2000ML. THIS EVENT MEETS OVERFILL CRITERIA.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HOMECHOICE CYCLER-REFURBISHED SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1