FDA Adverse Event Malfunction Summary report: N

6600

MDR report key: 1861206 · Received October 4, 2010

Report

Report Number
1720753-2010-03363
Event Type
Malfunction
Date Received
October 4, 2010
Date of Event
June 17, 2010
Report Date
October 4, 2010
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE REPRESENTATIVE EVALUATED THE SYSTEM AND REPLACED THE X-RAY CONTROLLER BOARD. SYSTEM OPERATES AS INTENDED.

Description of Event or Problem · 1

CUSTOMER REPORTED THE SYSTEM WILL LOCK UP DURING EXTENDED X-RAYS. NO PATIENT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 6600 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 6600

Patients

Seq Age Sex Outcome Treatment
1