CONTAK RENEWAL
Report
- Report Number
- 2124215-2010-15532
- Event Type
- Malfunction
- Date Received
- October 8, 2010
- Date of Event
- July 22, 2010
- Report Date
- August 17, 2010
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- NIK
- Removal / Correction Number
- Z-0949-05 THRU Z-0953-05
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
UPON RECEIPT AT BOSTON SCIENTIFIC'S POST MARKET QUALITY ASSURANCE LABORATORY, THE DEVICE HEADER WAS EXAMINED. VISUAL INSPECTION NOTED THAT A WRENCH TIP WAS BROKEN OFF IN THE UP POSITION IN THE RIGHT VENTRICULAR HEX SLOT. FURTHER MICROSCOPIC ANALYSIS NOTED THAT THE DISTAL ATRIAL SETSCREW WAS STUCK IN THE UP POSITION AND ITS RETAINER CAP WAS BENT. THE SETSCREWS WERE REMOVED FROM THE CONNECTOR BLOCK AND NO IRREGULARITIES IN THE SETSCREW THREADS OR IN THE THREADS OF THE CONNECTOR BLOCK WERE REVEALED. TECHNICIANS EXPOSED THE DEVICE TO SIMULATED HEART LOAD CONDITIONS, AND THEN TESTED THE PACING AND SENSING FUNCTIONS. THE DEVICE OPERATED APPROPRIATELY WITH NO INTERRUPTIONS IN THERAPY OUTPUT OR PROGRAMMER COMMUNICATION AT THE RETURNED PROGRAMMED SETTINGS. A SERIES OF DETAILED TESTS WERE ALSO PERFORMED TO MEASURE THE ELECTRICAL PERFORMANCE OF THE DEVICE. NORMAL FUNCTION WAS OBSERVED AND THE PACEMAKER DID NOT EXHIBIT ANY ANOMALIES. THE DAMAGE TO THE RIGHT ATRIAL RETAINER RING AND BROKEN TORQUE WRENCH TIP WAS INADVERTENTLY INDUCED.
UPON ANALYSIS THIS EVENT WILL BE UPDATED.
.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT DURING AN ELECTIVE REPLACEMENT OF THIS CARDIAC RESYCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) THE TORQUE WRENCH GOT STUCK INSIDE THE HEADER WHEN ATTEMPTING TO LOOSEN THE RIGHT VENTRICULAR LEAD SETSCREW. THE PHYSICIAN ELECTED TO CUT THE RIGHT VENTRICULAR (RV) LEAD PACE/SENSE PORTION AND IMPLANTED A COMPETITOR LEAD. DURING THE SAME EXPLANT PROCEDURE IT WAS ALSO NOT POSSIBLE TO LOOSEN THE SETSCREWS OF THE RIGHT ATRIAL LEAD PORT. THE RIGHT ATRIAL (RA) LEAD WAS ALSO CUT AND CAPPED. AT THIS TIME NO ATRIAL LEAD IS IMPLANTED. THE DEVICE WAS RETURNED FOR ANALYSIS. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CONTAK RENEWAL | IMPLANTABLE CHF GENERATOR | NIK | GUIDANT CRM CLONMEL IRELAND | H195 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 78 YR |