Manufacturer
INTUITIVE SURGICAL, INC
Product Code
GCJ
PMA / PMN Number
K191736
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

AN INVESTIGATION WAS COMPLETED TO DETERMINE THE CAUSE OF THIS REPORTED EVENT. THE REPORTED EVENT WAS ADDRESSED WITH PHONE SUPPORT. INTUITIVE SURGICAL, INC. (ISI) TECHNICAL SERVICE ENGINEER (TSE) ADVISED CUSTOMER TO USE BACKUP ENDOSCOPE TO RESOLVE REPORTED ISSUE. NO SITE VISIT WAS CONDUCTED. THE SYSTEM WAS WORKING PROPERLY, AND NO ADDITIONAL ACTION WAS REQUIRED AS THE ISSUE WAS RESOLVED.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED INCISIONAL HERNIA IPOM SURGICAL PROCEDURE, CUSTOMER CONTACTED INTUITIVE SURGICAL, INC. (ISI) TECHNICAL SERVICE ENGINEER (TSE) AND REPORTED DURING CASE IMAGE WAS UPSIDE DOWN. TSE REVIEWED LOGS AND OBSERVED ERROR #23003 OBSERVED IN LOGS ON UNIVERSAL SURGICAL MANIPULATOR (USM) 3 AXIS 4 WHERE ENDOSCOPE WAS INSTALLED. THE PROCEDURE WAS COMPLETING AS PLANNED WITH NO REPORTED INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
421072 NONE ENDOSCOPE PLUS GCJ INTUITIVE SURGICAL, INC 470057-08 N/A

Patients

Seq Age Sex Outcome Treatment
1 Unknown DA VINCI INSTRUMENTS AND ACCESSORIES