FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 1860979 · Received October 7, 2010

Report

Report Number
2649622-2010-10667
Event Type
Injury
Date Received
October 7, 2010
Date of Event
July 26, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S24
Removal / Correction Number
ASKU
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT T WAVE OVERSENSING WAS NOTED ALONG WITH NOISE ON THE FAR FIELD EGM AND SENSING INTEGRITY COUNTS. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6947 ASKU

Patients

Seq Age Sex Outcome Treatment
1 64 YR Required Intervention (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB