FDA Adverse Event Malfunction Summary report: N

ATTAIN ABILITY

MDR report key: 1860947 · Received October 7, 2010

Report

Report Number
2649622-2010-10549
Event Type
Malfunction
Date Received
October 7, 2010
Date of Event
July 19, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
OJX
PMA / PMN Number
P080006
Removal / Correction Number
ASKU
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4): THE FULL LEAD WAS RETURNED AND ANALYZED. PRIMARY ANALYSIS FINDING: NO ANOMALIES WERE FOUND. (B)(4): THE FULL LEAD WAS RETURNED AND ANALYZED. NO ANOMALIES WERE FOUND. BLOOD/BODY FLUID WAS PRESENT ON THE DISTAL CONDUCTOR.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN IMPLANT ATTEMPT, TWO DIFFERENT MODEL LEADS FAILED TO ACHIEVE ACCEPTABLE THRESHOLDS, NOR STABLE FIXATION DUE TO PATIENT ANATOMY. NEITHER LEAD WAS USED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ATTAIN ABILITY IMPLANTABLE PACING LEAD OJX MEDTRONIC PUERTO RICO, INC. 4196 ASKU

Patients

Seq Age Sex Outcome Treatment
1 72 YR Required Intervention