FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE WITH PRE-ATTACHED HOLDER

MDR report key: 18608934 · Received January 30, 2024

Report

Report Number
1024879-2024-00066
Event Type
Malfunction
Date Received
January 30, 2024
Date of Event
January 10, 2024
Report Date
February 7, 2024
Manufacturer
BECTON, DICKINSON & CO., (BD)
Product Code
FMI
UDI-DI
50382903686518
PMA / PMN Number
K982541
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE FOLLOWING FIELDS HAVE BEEN UPDATED WITH CORRECTED AND/OR ADDITIONAL INFORMATION. D10: DEVICE AVAILABLE FOR EVALUATION: YES. D10: RETURNED TO MANUFACTURER ON: 02-FEB-2023. H.6. INVESTIGATION SUMMARY: MATERIAL #: 368651. LOT/BATCH #: 3223395. BD RECEIVED 17 SAMPLES FOR INVESTIGATION. THE SAMPLES WERE EVALUATED BY FUNCTIONAL TESTING, EACH HAVING THEIR ECLIPSE SHIELD ACTIVATED, AND THE INDICATED FAILURE MODE FOR DEFECTIVE LOCKING MECHANISM WITH THE INCIDENT LOT WAS NOT OBSERVED AS ALL PRODUCT SPECIFICATIONS WERE MET. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT IS UNABLE TO BE CONFIRMED FOR THE INDICATED FAILURE MODE DEFECTIVE LOCKING MECHANISM. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE.

Additional Manufacturer Narrative · 0

H.3. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE USING BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE WITH PRE-ATTACHED HOLDER, THE SAFETY SHIELD FELL OFF OF THE DEVICE. NO PATIENT IMPACT REPORTED.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE USING BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE WITH PRE-ATTACHED HOLDER, THE SAFETY SHIELD FELL OFF OF THE DEVICE. NO PATIENT IMPACT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2538585 BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE WITH PRE-ATTACHED HOLDER HYPODERMIC SINGLE LUMEN NEEDLE FMI BECTON, DICKINSON & CO., (BD) 3223395 50382903686518

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown