FDA Adverse Event Malfunction Summary report: N

PD CYCLER 110 VOLT HOME CHOICEPRO AUTOMATED

MDR report key: 1860883 · Received October 7, 2010

Report

Report Number
1423500-2010-04138
Event Type
Malfunction
Date Received
October 7, 2010
Date of Event
August 8, 2010
Report Date
September 16, 2010
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
PMA / PMN Number
K053512
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE HAS BEEN RETURNED TO THE BAXTER PRODUCT ANALYSIS LAB FOR EVALUATION. UPON COMPLETION OF THE EVALUATION A FOLLOW UP REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). THE PRODUCT ANALYSIS LAB (PAL) EVALUATION AND HOMECHOICE (HC) DESIGN ENGINEERS AT DEKA REVIEW OF THIS INCIDENT CONCUR THAT NO DEVICE FAILURE OR MALFUNCTION WAS IDENTIFIED THAT COULD HAVE CONTRIBUTED TO THE 18 IIPVS FOUND IN THE DEVICE'S EVENT LOGS. THE DEVICE MET PERFORMANCE SPECIFICATION REQUIREMENTS RELATIVE TO THE ISSUES OF IIPV. THE PRODUCT ANALYSIS LAB (PAL) EVALUATION AND HOMECHOICE (HC) DESIGN ENGINEERS AT DEKA REVIEW OF THIS INCIDENT CONCUR THAT NO DEVICE FAILURE OR MALFUNCTION WAS IDENTIFIED THAT COULD HAVE CONTRIBUTED TO THE 18 IIPVS FOUND IN THE DEVICE'S EVENT LOGS. THE DEVICE MET PERFORMANCE SPECIFICATION REQUIREMENTS RELATIVE TO THE ISSUES OF IIPV. THE DEVICE HAS NOT BEEN PREVIOUSLY RETURNED FOR ANY ISSUES RELATED TO PEDIATRIC IIPV OR TIMEOUT DURING RITE. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4). THIS IS REPORT 15 OF 18 RELATED TO IIPV INCIDENTS FOUND IN THE DEVICE'S EVENT LOGS.

Description of Event or Problem · 1

REVISION SURGERY WAS REPORTED DUE TO LOOSENING. THE REVISION SURGERY WAS PERFORMED ON (B)(6) 2010.

Description of Event or Problem · 1

INTRAPERITONEAL VOLUME (IIPV) SITUATION WAS IDENTIFIED IN THE LOG OF A RETURNED HOMECHOICE (HC) DEVICE. THE IIPV OCCURRED ON (B)(6) 2010 DURING DRAIN CYCLE 2. THE PROGRAMMED FILL VOLUME WAS 140ML AND THE TOTAL DRAIN VOLUME WAS 219ML. THIS DRAIN VOLUME MEETS BAXTER'S IIPV CRITERIA. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS REPORTED. THIS EVENT IS THE FIFTEENTH OF 18 EVENTS OF IIPV NOTED IN THE PATIENT LOGS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PD CYCLER 110 VOLT HOME CHOICEPRO AUTOMATED SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1 14 MO THE SOLUTIONS THE PATIENT WAS PRESCRIBED FOR USE W