FDA Adverse Event
Malfunction
Summary report: N
NOVAMETRIX
MDR report key: 1860833
·
Received September 17, 2010
Report
- Report Number
- 1860833
- Event Type
- Malfunction
- Date Received
- September 17, 2010
- Date of Event
- September 17, 2010
- Report Date
- September 17, 2010
- Manufacturer
- GE HEALTHCARE
- Product Code
- CCK
- Product Problem
- Yes
- Report Source
- User Facility report
- Reporter Location
- WI, US
Narratives
Description of Event or Problem · 1
THE INSPIRED CO2 READINGS ARE DRIFTING ABOVE ZERO AT TIMES. TROUBLESHOOTING AT THE BEDSIDE VERIFIES THAT THE INSPIRED CO2 LEVEL SHOULD BE ZERO. WHEN THE SYSTEM IS RECALIBRATED IT WILL WORK FINE FOR ABOUT AN HOUR. NO PROBLEMS WITH THE HARDWARE HAVE BEEN FOUND BY CLINICAL ENGINEERING. GE IS STUDYING THE ISSUE WITH OUR STAFF.======================MANUFACTURER RESPONSE FOR END-TIDAL CO2 MONITOR======================THEY ARE COLLECTING INFORMATION AT THIS POINT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | NOVAMETRIX | ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE | CCK | GE HEALTHCARE | MSNCAPCO2MOD | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * |