ADVANTA 2 FRAME
Report
- Report Number
- 3006697241-2024-00005
- Event Type
- Malfunction
- Date Received
- January 30, 2024
- Date of Event
- January 11, 2024
- Report Date
- January 30, 2024
- Manufacturer
- HILL-ROM MEXICO
- Product Code
- FNJ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
THE BAXTER TECHNICIAN FOUND THE¿RIGHT CABLE AND CAREGIVER MEMBRANE NEEDED TO BE REPLACED. PER THE HILLROM SERVICE MANUAL THE ADVANTA¿ 2 SHOULD BE SUBJECT TO AN EFFECTIVE MAINTENANCE PROGRAM. THIS¿WILL HELP MAKE SURE OF A LONG, OPERATIVE LIFE FOR THE ADVANTA¿ 2 BED. THE PREVENTATIVE MAINTENANCE WILL HELP TO REDUCE DOWNTIME DUE TO EXCESSIVE WEAR FAILURES. AN ANNUAL SERVICE OF THE BED IS ADVISED IN ORDER TO MAINTAIN ITS CHARACTERISTICS AND PERFORMANCE. MAKE SURE THE HEAD AND FOOT SIDE RAILS ARE NOT BENT OR TWISTED. MAKE SURE THE LATCH MECHANISM OPERATES CORRECTLY. AN AUDIBLE CLICK MUST BE HEARD. REMOVE THE SIDE RAIL COVER, AND MAKE SURE THE MOUNTING SCREWS ARE TIGHT. INSPECT THE CABLE ROUTING FOR PINCHING, BINDING, AND DAMAGE. MAKE SURE ALL FUNCTIONS ON THE CAREGIVER CONTROL OPERATE CORRECTLY.REPAIR OR REPLACE THE SIDE RAIL AS NECESSARY. A SEARCH OF THE BAXTER MAINTENANCE RECORDS DID NOT SHOW BAXTER PERFORMED ANY PREVENTATIVE MAINTENANCE ON THIS BED. IT IS UNKNOWN IF THE FACILITY PERFORMS PREVENTATIVE MAINTENANCE ON THEIR BEDS. THE TECHNICIAN REPLACED THE RIGHT CABLE AND CAREGIVER MEMBRANE TO RESOLVE THE REPORTED EVENT. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.
BAXTER RECEIVED A REPORT FROM A BAXTER TECHNICIAN STATING THE BED FUNCTIONS SELF-RUN WHEN RAISING RIGHT SIDERAIL. THE BED WAS LOCATED AT THE ACCOUNT. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2538544 | ADVANTA 2 FRAME | BED, MANUAL | FNJ | HILL-ROM MEXICO | P1190A000109 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |