FDA Adverse Event Malfunction Summary report: N

ADVANTA 2 FRAME

MDR report key: 18608190 · Received January 30, 2024

Report

Report Number
3006697241-2024-00005
Event Type
Malfunction
Date Received
January 30, 2024
Date of Event
January 11, 2024
Report Date
January 30, 2024
Manufacturer
HILL-ROM MEXICO
Product Code
FNJ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THE BAXTER TECHNICIAN FOUND THE¿RIGHT CABLE AND CAREGIVER MEMBRANE NEEDED TO BE REPLACED. PER THE HILLROM SERVICE MANUAL THE ADVANTA¿ 2 SHOULD BE SUBJECT TO AN EFFECTIVE MAINTENANCE PROGRAM. THIS¿WILL HELP MAKE SURE OF A LONG, OPERATIVE LIFE FOR THE ADVANTA¿ 2 BED. THE PREVENTATIVE MAINTENANCE WILL HELP TO REDUCE DOWNTIME DUE TO EXCESSIVE WEAR FAILURES. AN ANNUAL SERVICE OF THE BED IS ADVISED IN ORDER TO MAINTAIN ITS CHARACTERISTICS AND PERFORMANCE. MAKE SURE THE HEAD AND FOOT SIDE RAILS ARE NOT BENT OR TWISTED. MAKE SURE THE LATCH MECHANISM OPERATES CORRECTLY. AN AUDIBLE CLICK MUST BE HEARD. REMOVE THE SIDE RAIL COVER, AND MAKE SURE THE MOUNTING SCREWS ARE TIGHT. INSPECT THE CABLE ROUTING FOR PINCHING, BINDING, AND DAMAGE. MAKE SURE ALL FUNCTIONS ON THE CAREGIVER CONTROL OPERATE CORRECTLY.REPAIR OR REPLACE THE SIDE RAIL AS NECESSARY. A SEARCH OF THE BAXTER MAINTENANCE RECORDS DID NOT SHOW BAXTER PERFORMED ANY PREVENTATIVE MAINTENANCE ON THIS BED. IT IS UNKNOWN IF THE FACILITY PERFORMS PREVENTATIVE MAINTENANCE ON THEIR BEDS. THE TECHNICIAN REPLACED THE RIGHT CABLE AND CAREGIVER MEMBRANE TO RESOLVE THE REPORTED EVENT. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 0

BAXTER RECEIVED A REPORT FROM A BAXTER TECHNICIAN STATING THE BED FUNCTIONS SELF-RUN WHEN RAISING RIGHT SIDERAIL. THE BED WAS LOCATED AT THE ACCOUNT. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2538544 ADVANTA 2 FRAME BED, MANUAL FNJ HILL-ROM MEXICO P1190A000109

Patients

Seq Age Sex Outcome Treatment
1 Unknown